Quality assurance by systematic in vivo dosimetry: results on a large cohort of patients
- PMID: 10869759
- DOI: 10.1016/s0167-8140(00)00195-x
Quality assurance by systematic in vivo dosimetry: results on a large cohort of patients
Abstract
Background: In vivo dosimetry is widely considered to be an important tool for quality assurance in external radiotherapy.
Introduction: In this study we report on our experience over more than 4 years in systematic in vivo dosimetry with diodes.
Materials and methods: From November '94 an in vivo entrance dosimetry check was performed for every new patient irradiated at one of our treatment units (Linac 6/100, 6 MV X-rays). Diodes were calibrated in terms of entrance dose; appropriate correction factors had been previously assessed (taking SSDs, field width, wedge, oblique incidence and blocking tray into account) and were individually applied to in vivo diode readings. The in vivo measured entrance dose was compared with the expected one, with a 5% action level; if a larger deviation was found, all treatment parameters were verified, and the in vivo dosimetry check was repeated. During the period November '94-May '99, 2824 measurements on 1433 patients were collected.
Results: Nine out of 1433 (0.63%) serious systematic errors (leading to a 5% or more on the delivered dose to the PTV) were detected by in vivo dosimetry; four out of nine would produce a 10% or more error if not detected. The rate of serious systematic errors detected by an independent check of treatment chart and MU calculation was found to be 1.5%, showing that less than 1/3 of the errors escapes this check. One hundred and twelve out of 1433 (7.8%) patients had more than one check: the rate of second checks was significantly higher for breast patients (31/250, 12.4%) against non-breast patients (81/1183, 6.8%, P=0.003). A number of patients demonstrated a persistent relatively large error even after two or more checks. For almost all patients the cause of the deviation was assessed; the most frequent cause was the difficulty in correctly positioning the patient and/or the diode. When analyzing the distribution of the deviations between measured and expected entrance doses (excluding first checks in the case of repetition of the in vivo dosimetry control) the mean deviation was 0.4% with a standard deviation equal to 3.0%. The rates of deviations larger than 5 and 7% were 9.9 and 2.6%, respectively. When considering the same data taking the average deviation in the case of opposed beams, the SD became 2.6% and the rates of deviations larger than 5 and 7%, respectively, 5.2 and 0.8%. When dividing the beams according to their orientation, significantly higher rates of large deviations (>5 and 7%) were found for oblique and posterior-anterior (PA) fields against lateral and anterior-posterior (AP) fields (P<0.05). Similarly, higher rates of large deviations were found for wedged fields against unwedged fields (P<0.03) and for blocked fields against unblocked fields (P<0.01). When dividing the data according to the anatomical district, accuracy was worse for breast (mean deviation 0.1%, 1 SD: 3.5%) and neck AP-PA fields (mean deviation 1%, 1 SD: 3,4%). Better accuracy was found for vertebrae (0.1%, 1 SD 2. 1%) and brain patients (-0.7%, 1 SD: 2.6%). During the considered period, in vivo dosimetry was also able to promptly detect a systematic error caused by a wrong resetting of the simulator height couch indicator, with a consequent error in the estimate of patient thickness of about 4 cm.
Conclusions: In our experience, systematic in vivo dosimetry demonstrated to be a valid tool for quality assurance, both in detecting systematic errors which may escape the data transfer/MU calculation check and in giving an effective way of estimating the accuracy of treatment delivery.
Similar articles
-
Quality assurance in radiation oncology. A study of feasibility and impact on action levels of an in vivo dosimetry program during breast cancer irradiation.Radiother Oncol. 1998 Apr;47(1):29-36. doi: 10.1016/s0167-8140(97)00202-8. Radiother Oncol. 1998. PMID: 9632289
-
In-vivo dosimetry by diode semiconductors in combination with portal films during TBI: reporting a 5-year clinical experience.Radiother Oncol. 1999 Sep;52(3):269-76. doi: 10.1016/s0167-8140(99)00104-8. Radiother Oncol. 1999. PMID: 10580875
-
In vivo dosimetry with diodes in a radiotherapy department in Pakistan.Radiat Prot Dosimetry. 2011 Nov;147(4):608-13. doi: 10.1093/rpd/ncq566. Epub 2011 Jan 2. Radiat Prot Dosimetry. 2011. PMID: 21199828
-
Is it necessary to repeat quality control procedures for head and neck patients?Radiother Oncol. 1991 Jul;21(3):201-10. doi: 10.1016/0167-8140(91)90038-i. Radiother Oncol. 1991. PMID: 1924856 Review.
-
In vivo dosimetry during external photon beam radiotherapy.Int J Radiat Oncol Biol Phys. 1999 Jan 15;43(2):245-59. doi: 10.1016/s0360-3016(98)00341-1. Int J Radiat Oncol Biol Phys. 1999. PMID: 10030247 Review.
Cited by
-
Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.Phys Imaging Radiat Oncol. 2022 May 13;22:85-90. doi: 10.1016/j.phro.2022.05.003. eCollection 2022 Apr. Phys Imaging Radiat Oncol. 2022. PMID: 35602547 Free PMC article.
-
Real-time dosimetry in external beam radiation therapy.World J Radiol. 2013 Oct 28;5(10):352-5. doi: 10.4329/wjr.v5.i10.352. World J Radiol. 2013. PMID: 24179630 Free PMC article.
-
A multi-institutional study of independent calculation verification in inhomogeneous media using a simple and effective method of heterogeneity correction integrated with the Clarkson method.J Radiat Res. 2018 Jul 1;59(4):490-500. doi: 10.1093/jrr/rry039. J Radiat Res. 2018. PMID: 29790997 Free PMC article.
-
In vivo dosimetry in UK external beam radiotherapy: current and future usage.Br J Radiol. 2017 Apr;90(1072):20160915. doi: 10.1259/bjr.20160915. Epub 2017 Feb 16. Br J Radiol. 2017. PMID: 28205452 Free PMC article.
-
Evaluation of target dose based on water-equivalent thickness in external beam radiotherapy.J Med Phys. 2013 Jan;38(1):44-51. doi: 10.4103/0971-6203.106605. J Med Phys. 2013. PMID: 23532059 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials