Sensitivity of Swedish interactive threshold algorithm compared with standard full threshold algorithm in Humphrey visual field testing
- PMID: 10889102
- DOI: 10.1016/s0161-6420(00)00140-8
Sensitivity of Swedish interactive threshold algorithm compared with standard full threshold algorithm in Humphrey visual field testing
Abstract
Objective: To compare the sensitivity of Swedish interactive threshold algorithm (SITA) strategies with the standard full threshold algorithm in the Humphrey Field Analyzer.
Design: Observational case series.
Participants: Forty-eight glaucoma patients who were experienced in automated perimetry.
Testing: Central field testing was performed with the 30-2 program using standard full threshold (SFT), SITA standard (SS), and SITA fast (SF) strategies. All three tests were carried out on each of four different days in a span of 4 weeks.
Main outcome measures: Sensitivity, repeatability, time saved, and the extent of defect in the SITA strategies were compared with those of the SFT.
Results: The sensitivity of SS and SF were 95.12% and 92.68%, respectively. The time saved in SS and SF was 53.12+/-9.51% and 70.69+/-8.81%, respectively. The repeatability as assessed by intraclass correlation showed excellent repeatability for the SFT and SS strategies and excellent to poor repeatability with the SF strategy. With increasing mean deviation, the defects (significant at P<0.5%) in the pattern deviation plots tended to be more in the SITA strategies as compared with SFT.
Conclusions: Swedish interactive threshold algorithm strategies have good sensitivity and are significantly faster as compared with the standard threshold algorithm. The repeatability of the SFT and SS strategies are excellent, whereas that of the SF strategy is variable.
Similar articles
-
Sensitivity and specificity of the Swedish interactive threshold algorithm for glaucomatous visual field defects.Ophthalmology. 2002 Jun;109(6):1052-8. doi: 10.1016/s0161-6420(02)01047-3. Ophthalmology. 2002. PMID: 12045043
-
Comparison of glaucomatous visual field defects using standard full threshold and Swedish interactive threshold algorithms.Arch Ophthalmol. 2002 Sep;120(9):1136-41. doi: 10.1001/archopht.120.9.1136. Arch Ophthalmol. 2002. PMID: 12215086
-
Swedish Interactive Threshold Algorithm for central visual field defects unrelated to nerve fiber layer.Graefes Arch Clin Exp Ophthalmol. 2016 May;254(5):845-54. doi: 10.1007/s00417-015-3132-x. Epub 2015 Aug 18. Graefes Arch Clin Exp Ophthalmol. 2016. PMID: 26279004
-
Visual field testing in glaucoma using the Swedish Interactive Thresholding Algorithm (SITA).Surv Ophthalmol. 2025 Jan-Feb;70(1):141-152. doi: 10.1016/j.survophthal.2024.09.005. Epub 2024 Sep 29. Surv Ophthalmol. 2025. PMID: 39349186 Review.
-
Current practice with standard automated perimetry.Semin Ophthalmol. 2000 Dec;15(4):172-81. doi: 10.3109/08820530009037869. Semin Ophthalmol. 2000. PMID: 17585432 Review.
Cited by
-
Comparison of different analytic algorithms for interpretation of the Swedish interactive threshold algorithm strategy.Clinics (Sao Paulo). 2008 Jun;63(3):333-8. doi: 10.1590/s1807-59322008000300008. Clinics (Sao Paulo). 2008. PMID: 18568242 Free PMC article.
-
Fractional anisotropy of the optic radiations correlates with the visual field after epilepsy surgery.Neuroradiology. 2019 Dec;61(12):1425-1436. doi: 10.1007/s00234-019-02281-2. Epub 2019 Sep 7. Neuroradiology. 2019. PMID: 31494681
-
Early isolated optic neuropathy caused by cyclosporine.Int Ophthalmol. 2019 Apr;39(4):853-860. doi: 10.1007/s10792-018-0886-2. Epub 2018 Mar 12. Int Ophthalmol. 2019. PMID: 29532216
-
Nomograms for Converting Perimetric Sensitivity From Full Threshold and SITA Fast to SITA Standard in Patients With Glaucoma and Healthy Subjects.Transl Vis Sci Technol. 2021 Aug 2;10(9):2. doi: 10.1167/tvst.10.9.2. Transl Vis Sci Technol. 2021. PMID: 34342610 Free PMC article.
-
Reversal of Glaucoma Hemifield Test Results and Visual Field Features in Glaucoma.Ophthalmology. 2018 Mar;125(3):352-360. doi: 10.1016/j.ophtha.2017.09.021. Epub 2017 Nov 2. Ophthalmology. 2018. PMID: 29103791 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical