A noninvasive method of measuring wave intensity, a new hemodynamic index: application to the carotid artery in patients with mitral regurgitation before and after surgery
- PMID: 10901480
- DOI: 10.1007/BF03257237
A noninvasive method of measuring wave intensity, a new hemodynamic index: application to the carotid artery in patients with mitral regurgitation before and after surgery
Abstract
Wave intensity (WI) is a new hemodynamic index, which is defined as (dP/dt)(dU/dt) at any site of the circulation, where dP/dt and dU/dt are the time derivatives of blood pressure and velocity, respectively. Arterial WI in normal subjects has two positive sharp peaks. The first peak occurs during early systole when a forward-traveling compression wave is generated by the left ventricle. The magnitude of this peak increases markedly with an increase in cardiac contractility. The second peak, which occurs towards the end of systole, is caused by generation of a forward-traveling expansion wave by the ability of the left ventricle to actively stop aortic blood flow. The interval between the R wave of the ECG and the first peak of WI (R-1st peak interval) and the interval between the first and second peaks (1st-2nd interval) are approximately equal to the preejection period and left ventricular ejection time, respectively. Using a combined Doppler and echo-tracking system, we obtained carotid arterial WI noninvasively. We examined the characteristics of WI in 11 patients with mitral regurgitation (MR) before and after surgery, and 24 normal volunteers. In the MR group before surgery, the second peak was decreased and the (1st-2nd interval)/(R-R interval) ratio was reduced, compared with the normal group (140 +/- 130 vs 750 +/- 290mmHg m/s3. P < 0.0083; 20.7% +/- 3.4% vs 26.7% +/- 2.8%, P < 0.083). There were no significant differences in the first peak between the normal group and the MR group before and after surgery. The second peak in the MR group was increased significantly (P < 0.016 vs before surgery) to 1,150 +/- 830mmHg m/s3 in the early period after surgery (stage I), and to 1,090 +/- 580mmHgm/s3 in the late period after surgery (stage II). These values did not differ significantly from that of the normal group. At stage I, the (R-1st peak interval)/ (R-R interval) ratio was increased from 13.4% +/- 2.7% to 20.6% +/- 5.6% (P < 0.016 vs before surgery). At stage II, this ratio decreased to 16.2% +/- 2.8% (P < 0.016 vs stage I). but was still significantly higher than that before surgery. The (1st-2nd interval)/(R-R interval) ratio increased significantly after surgery (P < 0.016 vs before surgery) to values (27.0% +/- 4.5% at stage I and 28.9% +/- 2.6% at stage II) which did not differ significantly from that of the normal group. The recovery of the second peak after surgery suggests that the left ventricle had recovered the ability to actively stop aortic blood flow. Wave intensity is useful for analyzing changes in the working condition of the heart.
Similar articles
-
Clinical usefulness of carotid arterial wave intensity in assessing left ventricular systolic and early diastolic performance.Heart Vessels. 2003 Jul;18(3):107-11. doi: 10.1007/s00380-003-0700-5. Heart Vessels. 2003. PMID: 12955424
-
A new noninvasive measurement system for wave intensity: evaluation of carotid arterial wave intensity and reproducibility.Heart Vessels. 2002 Nov;17(1):12-21. doi: 10.1007/s003800200037. Heart Vessels. 2002. PMID: 12434197
-
Early Hemodynamic Improvement after Percutaneous Mitral Valve Repair Evaluated by Noninvasive Pressure-Volume Analysis.J Am Soc Echocardiogr. 2016 Sep;29(9):888-98. doi: 10.1016/j.echo.2016.05.012. Epub 2016 Jun 29. J Am Soc Echocardiogr. 2016. PMID: 27372560 Clinical Trial.
-
Associations of increased arterial stiffness with left ventricular ejection performance and right ventricular systolic pressure in mitral regurgitation before and after surgery: Wave intensity analysis.Int J Cardiol Heart Vasc. 2017 Jul 12;16:7-13. doi: 10.1016/j.ijcha.2017.06.002. eCollection 2017 Sep. Int J Cardiol Heart Vasc. 2017. PMID: 29067354 Free PMC article.
-
Afterload mismatch in aortic and mitral valve disease: implications for surgical therapy.J Am Coll Cardiol. 1985 Apr;5(4):811-26. doi: 10.1016/s0735-1097(85)80418-6. J Am Coll Cardiol. 1985. PMID: 3882814 Review.
Cited by
-
Clinical usefulness of wave intensity analysis.Med Biol Eng Comput. 2009 Feb;47(2):197-206. doi: 10.1007/s11517-008-0388-x. Epub 2008 Sep 2. Med Biol Eng Comput. 2009. PMID: 18763005
-
Study of the Characteristics of Pulmonary Trunk in Pulmonary Hypertension Secondary to Left Heart Disease Using Pressure-Velocity Loops (PU-Loops).Ann Vasc Dis. 2017 Sep 25;10(3):197-204. doi: 10.3400/avd.oa.17-00064. Ann Vasc Dis. 2017. PMID: 29147168 Free PMC article.
-
Effects of sublingual nitroglycerin on working conditions of the heart and arterial system: analysis using wave intensity.J Med Ultrason (2001). 2005 Dec;32(4):145-52. doi: 10.1007/s10396-005-0057-8. J Med Ultrason (2001). 2005. PMID: 27277481
-
Wave intensity analysis of carotid artery: a noninvasive technique for assessing hemodynamic changes of hyperthyroid patients.J Huazhong Univ Sci Technolog Med Sci. 2010 Oct;30(5):672-7. doi: 10.1007/s11596-010-0563-9. Epub 2010 Nov 10. J Huazhong Univ Sci Technolog Med Sci. 2010. PMID: 21063855
-
Wave intensity as a useful modality for assessing ventilation-perfusion imbalance in subclinical patients with hypertension.Heart Vessels. 2018 Aug;33(8):931-938. doi: 10.1007/s00380-018-1138-0. Epub 2018 Feb 12. Heart Vessels. 2018. PMID: 29435641
References
Publication types
MeSH terms
LinkOut - more resources
Other Literature Sources