Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing
- PMID: 10923072
- DOI: 10.5858/2000-124-1135-EODHLO
Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing
Abstract
Objectives: To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors.
Design: Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared.
Setting: Six hand-held glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL).
Results: Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration.
Conclusions: We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.
Comment in
-
Connectivity. The millennium challenge for point-of-care testing.Arch Pathol Lab Med. 2000 Aug;124(8):1108-10. doi: 10.5858/2000-124-1108-C. Arch Pathol Lab Med. 2000. PMID: 10923064 No abstract available.
Similar articles
-
Can one point-of-care glucose meter be used for all pediatric and adult hospital patients? Evaluation of three meters, including recently modified test strips.Diabetes Technol Ther. 2011 Jan;13(1):55-62. doi: 10.1089/dia.2010.0129. Diabetes Technol Ther. 2011. PMID: 21175272
-
Plasma-equivalent glucose at the point-of-care: evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters.Clin Chim Acta. 2007 Nov-Dec;386(1-2):63-8. doi: 10.1016/j.cca.2007.07.025. Epub 2007 Aug 7. Clin Chim Acta. 2007. PMID: 17727831
-
Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.Crit Care Med. 2001 May;29(5):1062-70. doi: 10.1097/00003246-200105000-00038. Crit Care Med. 2001. PMID: 11378622 Clinical Trial.
-
A glucose meter evaluation co-designed with both health professional and consumer input.N Z Med J. 2013 Nov 22;126(1386):90-7. N Z Med J. 2013. PMID: 24316996 Review.
-
Comparison of Accuracy Guidelines for Hospital Glucose Meters.J Diabetes Sci Technol. 2020 May;14(3):546-552. doi: 10.1177/1932296819898277. Epub 2020 Jan 25. J Diabetes Sci Technol. 2020. PMID: 31983225 Free PMC article. Review.
Cited by
-
Pseudohyperglycaemia in a comatose patient after picking cherries.BMJ Case Rep. 2016 Oct 28;2016:bcr2016218141. doi: 10.1136/bcr-2016-218141. BMJ Case Rep. 2016. PMID: 27793876 Free PMC article.
-
Approaches for improving glucose monitor measurements for self-monitoring of blood glucose: from measurement harmonization to external quality assessment programs.J Diabetes Sci Technol. 2007 Mar;1(2):153-7. doi: 10.1177/193229680700100202. J Diabetes Sci Technol. 2007. PMID: 19888399 Free PMC article.
-
Pre-/analytical factors affecting whole blood and plasma glucose concentrations in loggerhead sea turtles (Caretta caretta).PLoS One. 2020 Mar 3;15(3):e0229800. doi: 10.1371/journal.pone.0229800. eCollection 2020. PLoS One. 2020. PMID: 32126109 Free PMC article.
-
Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curves.Clin Chim Acta. 2008 Mar;389(1-2):31-9. doi: 10.1016/j.cca.2007.11.019. Epub 2007 Dec 3. Clin Chim Acta. 2008. PMID: 18157943 Free PMC article.
-
Assessing the performance of handheld glucose testing for critical care.Diabetes Technol Ther. 2008 Dec;10(6):445-51. doi: 10.1089/dia.2008.0049. Diabetes Technol Ther. 2008. PMID: 19049373 Free PMC article.