Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Case Reports
. 2000 Aug;46(8 Pt 1):1157-61.

An IgM lambda antibody to Escherichia coli produces false-positive results in multiple immunometric assays

Affiliations
  • PMID: 10926897
Case Reports

An IgM lambda antibody to Escherichia coli produces false-positive results in multiple immunometric assays

M Covinsky et al. Clin Chem. 2000 Aug.

Abstract

Background: Interferences in immunometric assays as a result of human anti-immunoglobulin antibodies frequently have been described in the literature. The etiology of these interfering antibodies is usually not known but has been associated with rheumatoid factors in some assays. It is known that microorganisms in experimental settings can induce anti-immunoglobulin antibodies.

Methods: Following Escherichia coli septicemia, a 56-year-old male patient had increased immunoassay results for cardiac troponin I, thyrotropin, human chorionic gonadotropin, alpha-fetoprotein, and CA-125 that were consistent with myocardial infarction, hyperthyroidism, and pregnancy, and suggestive of an occult neoplasm such as hepatic or ovarian cancer. None of these diagnoses were consistent with the rest of his medical exam. In addition, the patient had a restricted IgM lambda paraprotein by immunofixation. Plasma from the patient was incubated with Sepharose-conjugated protein A, irrelevant murine monoclonal antibodies, and formalin-killed E. coli organisms from his infection to determine whether these immunoassay values were falsely increased.

Results: Incubation of the patient's plasma with irrelevant murine monoclonal antibodies or the E. coli organism produced normal immunoassay values and removed the IgM lambda paraprotein.

Conclusions: The patient produced a very restricted IgM lambda antibody response to the E. coli infection that had anti-immunoglobulin activity and caused falsely increased values in numerous immunometric assays. Microorganism-induced anti-immunoglobulin antibodies are discussed in the context of this patient.