Inadvertent corneal indentation can cause artifactitious widening of the iridocorneal angle on ultrasound biomicroscopy

Affiliations
  • PMID: 10928676
Case Reports

Inadvertent corneal indentation can cause artifactitious widening of the iridocorneal angle on ultrasound biomicroscopy

H Ishikawa et al. Ophthalmic Surg Lasers. 2000 Jul-Aug.

Abstract

A 70-year-old Asian female with narrow angles underwent ultrasound biomicroscopy (UBM) darkroom provocative testing. The angle was narrow under light conditions and occluded in the dark, except for the inferior quadrant where the angle was wide and the iris concave. This was reproducible in four quadrants with the use of a small UBM eye cup but not the large one. A small eye cup can indent the cornea, resulting in artifactitious angle widenening. Care must be taken when using the small eye cup not to press on the cornea.