CO assessment by suprasternal Doppler in critically ill patients: comparison with thermodilution
- PMID: 10945385
- DOI: 10.1007/s001340051234
CO assessment by suprasternal Doppler in critically ill patients: comparison with thermodilution
Abstract
Objective: Comparison of suprasternal Doppler (SST) and thermodilution (TD) for the measurement of cardiac output (CO) in critically ill patients.
Design: Prospective study.
Setting: Intensive care unit of a university hospital.
Patients and participants: 65 consecutive critically ill patients requiring a pulmonary artery catheter.
Interventions: Paired CO measurements were made simultaneously using SST and TD by two independent operators. The time to obtain a CO value by SST was measured. Correlation coefficients and the linear regression equation were determined. A Bland and Altman diagram was plotted. A Bland and Altman diagram was also plotted for the level of cardiac index (CI) values (low: CI < 2.5 l min(-1) m(-2); normal: 2.5 < or = CI < or = 4.5 l min(-1) m(-2); high: CI > 4.5 l min(-1) m(-2)).
Measurements and results: In seven patients SST failed to measure CO. In the remaining 58 patients 314 paired CO measurements were performed. The mean time to measure CO by SST was 73 +/- 45 s. The equation of linear regression was: SST(CO) = 0.84 TD(CO) + 1.39. The correlation coefficient was 0.84. The bias between SST and TD was -0.2 +/- 1.4 l min(-1). Biases were -0.23 +/- 0.50, -0.20 +/- 0.68, and 0.25 +/- 0.92 l min(-1) m(-2) for low, normal, and high levels of CI, respectively.
Conclusion: SST does not accurately measure CO but allows a rapid assessment of CI level in critically ill patients.
Similar articles
-
Noninvasive cardiac output monitoring by aortic blood flow determination: evaluation of the Sometec Dynemo-3000 system.Crit Care Med. 1998 Dec;26(12):2066-72. doi: 10.1097/00003246-199812000-00043. Crit Care Med. 1998. PMID: 9875922
-
Cardiac output measurement in critically ill patients: comparison of continuous and conventional thermodilution techniques.Can J Anaesth. 1995 Nov;42(11):972-6. doi: 10.1007/BF03011067. Can J Anaesth. 1995. PMID: 8590506
-
Training is required to improve the reliability of esophageal Doppler to measure cardiac output in critically ill patients.Intensive Care Med. 1998 Apr;24(4):347-52. doi: 10.1007/s001340050578. Intensive Care Med. 1998. PMID: 9609413
-
The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults.Intensive Care Med. 2004 Nov;30(11):2060-6. doi: 10.1007/s00134-004-2430-2. Epub 2004 Sep 11. Intensive Care Med. 2004. PMID: 15368037 Review.
-
Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.PLoS One. 2019 Oct 3;14(10):e0222105. doi: 10.1371/journal.pone.0222105. eCollection 2019. PLoS One. 2019. PMID: 31581196 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous