Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2000 Jul-Aug;7(4):301-11.
doi: 10.1067/mnc.2000.105279.

Accuracy of the automated assessment of left ventricular function with gated perfusion SPECT in the presence of perfusion defects and left ventricular dysfunction: correlation with equilibrium radionuclide ventriculography and echocardiography

Affiliations
Comparative Study

Accuracy of the automated assessment of left ventricular function with gated perfusion SPECT in the presence of perfusion defects and left ventricular dysfunction: correlation with equilibrium radionuclide ventriculography and echocardiography

T Chua et al. J Nucl Cardiol. 2000 Jul-Aug.

Abstract

Background: Gated single photon emission computed tomography (SPECT) with automated methods allows the quantitative assessment of left ventricular function and perfusion; however, its accuracy must be defined for patients with large earlier infarctions and severe rest perfusion defects, in whom the estimation of endocardial and epicardial borders might be more difficult, even with automated edge-detection techniques.

Methods and results: We prospectively compared the automated measurements of left ventricular ejection fraction (LVEF) and volumes from rest-injected gated Technetium 99m (Tc99m) perfusion SPECT with equilibrium radionuclide angiocardiography (ERNA) in 62 patients and the assessment of regional function with echocardiography in 22 patients. Forty-six patients had an earlier myocardial infarction (mean defect size, 34% of left ventricle; SD, 12.7%; range, 8% to 56%); 27 patients had large defects (> or = 20% of left ventricle; LVEF range, 8% to 75%). LVEF, as determined with Cedars-Sinai software (quantitative gated SPECT), correlated well with ERNA (r = 0.941; y = 1.003x + 1.15; P<.0001; SE of the estimate = 6.3%; mean difference -1.3% for LVEF) in the entire study population and in the subgroups of patients with an earlier infarction, severe defects, and large infarctions (> or = 20% of the left ventricle). A correlation existed between gated SPECT and ERNA volumes (r = 0.882, y = 1.040x - 14.7, P<.0001 for end-diastolic volume; r = 0.954, y = 1.147x - 13.9, P<.0001 for end-systolic volumes with the count-ratio technique), but with wider limits of agreement. The exact segmental score agreement between gated SPECT and echocardiography for regional function was 79.8% (281 of 352, kappa = 0.682).

Conclusions: Automated gated SPECT provides an accurate assessment of ejection fraction and regional function, even in the presence of an earlier myocardial infarction with large perfusion defects and significant left ventricular dysfunction.

PubMed Disclaimer

References

    1. J Am Coll Cardiol. 1997 Nov 1;30(5):1360-7 - PubMed
    1. J Nucl Med. 1999 May;40(5):805-9 - PubMed
    1. Eur J Nucl Med. 1996 Aug;23(8):909-16 - PubMed
    1. Mayo Clin Proc. 1986 Apr;61(4):254-62 - PubMed
    1. J Nucl Cardiol. 1998 Jul-Aug;5(4):387-94 - PubMed

Publication types

Substances

LinkOut - more resources