[Tissue harmonic imaging in comparison with conventional sonography: effect on image quality and observer variability in the measurement of the intima-media thickness in the common carotid artery]
- PMID: 10962992
- DOI: 10.1055/s-2000-4653
[Tissue harmonic imaging in comparison with conventional sonography: effect on image quality and observer variability in the measurement of the intima-media thickness in the common carotid artery]
Abstract
Objective: The purpose of this study was to determine if tissue harmonic imaging (THI) produced a higher image quality than conventional sonography and if THI reduced the observer availability in the intima media thickness (IMT) measurements of the common carotid artery (ACC).
Subjects and methods: A prospective study was performed on 26 healthy volunteers and IMT was assessed by three experienced sonographers in 52 carotid arteries. A 7.5 MHz transducer was used to measure carotid IMT in both harmonic and fundamental modes. Three measurements of IMT were carried out on the far wall of the ACC along a 1 cm-long section proximal to the bifurcation. Image quality was graded from 1-3 (1 = excellent, 2 = good, 3 = poor). IMT measurements using harmonic and fundamental modes were compared and intra- and inter-observer variability were calculated. Statistical analyses were performed using a kappa analysis, ANOVA and the McNemar/Wilcoxon Test.
Results: In the 52 cases image quality was excellent in 51% using THI in comparison to 32% using conventional sonography. Image quality was graded significantly more frequent as excellent using harmonic sonography as compared to conventional sonography (p = 0.001). A significant difference in IMT was found between the measurements obtained by harmonic imaging in comparison to those obtained by conventional imaging (p < 0.0001). Using the harmonic mode inter-observer variability of the IMT measurements was significantly reduced from 8.1% to 5.2% (p < 0.001).
Conclusions: THI produced a higher image quality with a better visualisation of carotid IMT and reduced significantly the intra- and inter-observer variability in the IMT measurements.
Similar articles
-
Reproducibility of sonographic measurements of carotid intima-media thickness.Acta Radiol. 2008 Dec;49(10):1162-6. doi: 10.1080/02841850802438520. Acta Radiol. 2008. PMID: 18846454
-
Multi-examiner reliability of automated radio frequency-based ultrasound measurements of common carotid intima-media thickness in rheumatoid arthritis.Rheumatology (Oxford). 2011 Oct;50(10):1860-4. doi: 10.1093/rheumatology/ker206. Epub 2011 Jun 29. Rheumatology (Oxford). 2011. PMID: 21719420
-
[Determination of intima-media thickness of the carotid artery: influences of methods, proband and examination variables].Ultraschall Med. 1998 Aug;19(4):168-73. doi: 10.1055/s-2007-1000484. Ultraschall Med. 1998. PMID: 9816621 German.
-
[Intima media thickness of the carotid arteries: early pointer to arteriosclerosis and therapeutic endpoint].Ultraschall Med. 2003 Jun;24(3):162-74. doi: 10.1055/s-2003-40058. Ultraschall Med. 2003. PMID: 12817310 Review. German.
-
Reproducibility of in vivo carotid intima-media thickness measurements: a review.Stroke. 1997 Mar;28(3):665-71. doi: 10.1161/01.str.28.3.665. Stroke. 1997. PMID: 9056629 Review.
Cited by
-
[Modern gray-scale sonography of the breast].Radiologe. 2010 Nov;50(11):964, 966-72. doi: 10.1007/s00117-010-2007-8. Radiologe. 2010. PMID: 20890590 Review. German.
-
[Latest developments in ultrasound of the liver].Radiologe. 2011 Aug;51(8):661-70. doi: 10.1007/s00117-010-2124-4. Radiologe. 2011. PMID: 21847777 Review. German.
-
Medical ultrasound systems.Interface Focus. 2011 Aug 6;1(4):477-89. doi: 10.1098/rsfs.2011.0027. Epub 2011 May 18. Interface Focus. 2011. PMID: 22866226 Free PMC article.
-
[Modern ultrasound diagnostics of deep vein thrombosis in lung embolism of unknown origin].Radiologe. 2007 Aug;47(8):673-84. doi: 10.1007/s00117-007-1530-8. Radiologe. 2007. PMID: 17634909 German.
-
[Current developments in vascular ultrasound].Radiologe. 2009 Nov;49(11):998-1004. doi: 10.1007/s00117-009-1874-3. Radiologe. 2009. PMID: 19859686 German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources