Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2000 Sep 1;48(2):435-42.
doi: 10.1016/s0360-3016(00)00625-8.

Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor

Affiliations

Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor

H Shirato et al. Int J Radiat Oncol Biol Phys. .

Abstract

Purpose: To achieve precise three-dimensional (3D) conformal radiotherapy for mobile tumors, a new radiotherapy system and its treatment planning system were developed and used for clinical practice.

Methods and materials: We developed a linear accelerator synchronized with a fluoroscopic real-time tumor tracking system by which 3D coordinates of a 2.0-mm gold marker in the tumor can be determined every 0.03 second. The 3D relationships between the marker and the tumor at different respiratory phases are evaluated using CT image at each respiratory phase, whereby the optimum phase can be selected to synchronize with irradiation (4D treatment planning). The linac is triggered to irradiate the tumor only when the marker is located within the region of the planned coordinates relative to the isocenter.

Results: The coordinates of the marker were detected with an accuracy of +/- 1 mm during radiotherapy in the phantom experiment. The time delay between recognition of the marker position and the start or stop of megavoltage X-ray irradiation was 0.03 second. Fourteen patients with various tumors were treated by conformal radiotherapy with a "tight" planning target volume (PTV) margin. They were surviving without relapse or complications with a median follow-up of 6 months.

Conclusion: Fluoroscopic real-time tumor tracking radiotherapy following 4D treatment planning was developed and shown to be feasible to improve the accuracy of the radiotherapy for mobile tumors.