Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters

Affiliations

Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters

R A Omary et al. J Vasc Interv Radiol. 2000 Sep.

Abstract

Purpose: To test the hypothesis that real-time magnetic resonance (MR) imaging-guided passive catheter tracking is feasible with use of dilute gadolinium (Gd)-filled catheters, to determine the optimal Gd concentration required for tracking, and to measure catheter tip tracking accuracy.

Materials and methods: The authors tested a real-time, T1-weighted, two-dimensional, spoiled gradient-recalled echo MR imaging sequence suitable for tracking catheters. In a yogurt phantom, the authors placed 5-F catheters filled with 2%-12% Gd solutions. MR imaging was performed with and without use of a projection dephaser that suppressed background signal. The authors measured signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and enhancement ratio to determine the optimal Gd concentration for catheter depiction. Catheter tip tracking accuracy was measured in an acrylic phantom with use of linear regression analysis, with goodness of fit assessed statistically with the F test.

Results: Peak catheter SNR, CNR, and enhancement ratios were obtained with 4%-6% Gd concentrations. Tip tracking accuracy was determined to be +/- 0.41 mm (R2 = 0.99; P < .0001). MR imaging reconstructions were displayed up to 3.1 frames/sec.

Conclusions: Accurate MR imaging-guided passive catheter tracking was feasible in real-time with use of dilute Gd-filled catheters. This technique may have application in MR imaging-guided endovascular procedures.