[Whole-body MRA on a rolling table platform (AngioSURF)]
- PMID: 11013607
- DOI: 10.1055/s-2000-7177
[Whole-body MRA on a rolling table platform (AngioSURF)]
Abstract
Purpose: Development of a technique for whole-body MR angiography based on a rolling table platform and integration of a surface coil.
Material and methods: The developed rolling table platform AngioSURF (System for Unlimited Rolling Field-of-view) with integrated surface coil can be mounted on top of the original patient table of a Siemens Symphony System. Data acquisition was performed with a standard body array surface coil. The system was tested on three volunteers and one patient with angiographically documented vascular pathology. Data acquisition was performed with a 3D-FLASH-sequence (TR/TE 2.1/0.7 ms, flip angle: 20 degrees, FOV 40 x 40 cm, 80 partitions, matrix 512 x 420 with zero interpolation). Five data sets were collected in immediate succession during continuous injection of a paramagnetic contrast agent. Time of acquisition per data set was 10 seconds. Table repositioning was performed manually within 3 seconds. Thus the total acquisition time amounted to 72 seconds.
Results: No problems with handling occurred in any of the four cases. The excellent image quality enables detailed assessment of the displayed vascular territories.
Conclusions: The rolling table platform with integrated surface coil (AngioSURF) allows diagnostic display of the arterial vascular system from supraaortic vessels to the distal trifurcation arteries in only 72 seconds.
Similar articles
-
High spatial resolution whole-body MR angiography featuring parallel imaging: initial experience.Rofo. 2004 Feb;176(2):163-9. doi: 10.1055/s-2004-817623. Rofo. 2004. PMID: 14872368
-
[Contrast media-enhanced 3D MR angiography of the peripheral vessels using a "tracking technique": preliminary results].Rofo. 1998 Jan;168(1):90-4. doi: 10.1055/s-2007-1015187. Rofo. 1998. PMID: 9501940 German.
-
[Magnetic resonance angiography of the vessels of the neck: the optimization of a dynamic technic during the rapid infusion of a paramagnetic contrast medium].Radiol Med. 1997 Oct;94(4):325-8. Radiol Med. 1997. PMID: 9465238 Italian.
-
[Diagnosis of renal artery stenosis with magnetic resonance angiography and stenosis quantification].J Mal Vasc. 2000 Dec;25(5):312-320. J Mal Vasc. 2000. PMID: 11148391 Review. French.
-
[MR Angiography: First choice for diagnosis of the arterial vascular system].Rofo. 2002 May;174(5):551-61. doi: 10.1055/s-2002-28317. Rofo. 2002. PMID: 11997853 Review. German.
Cited by
-
Whole-body magnetic resonance imaging: technique, guidelines and key applications.Ecancermedicalscience. 2021 Jan 7;15:1164. doi: 10.3332/ecancer.2021.1164. eCollection 2021. Ecancermedicalscience. 2021. PMID: 33680078 Free PMC article. Review.
-
[Whole body MRI--diagnostic strategy of the future?].Radiologe. 2007 Oct;47(10):904-14. doi: 10.1007/s00117-006-1411-6. Radiologe. 2007. PMID: 16998659 Review. German.
-
[Oncologic screening with whole-body MRI: possibilities and limitations].Radiologe. 2004 Sep;44(9):854-63. doi: 10.1007/s00117-004-1104-y. Radiologe. 2004. PMID: 15349733 Review. German.
-
[Cardiovascular whole-body MRI: possibilities and limitations in prevention].Radiologe. 2004 Sep;44(9):826-34. doi: 10.1007/s00117-004-1099-4. Radiologe. 2004. PMID: 15349731 German.
-
Whole body MR angiography screening.Int J Cardiovasc Imaging. 2004 Dec;20(6):587-91. doi: 10.1007/s10554-004-7033-z. Int J Cardiovasc Imaging. 2004. PMID: 15856646
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical