Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: a manifestation of T wave oversensing by ICDs
- PMID: 11025907
- DOI: 10.1111/j.1540-8159.2000.tb00980.x
Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: a manifestation of T wave oversensing by ICDs
Abstract
Preventive pacing algorithms designed to eliminate the pause that follows a premature ventricular depolarization have been incorporated in current implantable defibrillators. We report a patient in whom intermittent T wave oversensing frequently invoked the ventricular rate stabilization algorithm and resulted in periods of inappropriate VVI pacing. The problem was solved by decreasing the maximum sensitivity from 0.3 mV to 0.45 mV. Implant testing had revealed adequate sensing of ventricular fibrillation with a maximum sensitivity of 1.2 mV.
Similar articles
-
Ventricular rate stabilization for treatment of recurrent VT.Herzschrittmacherther Elektrophysiol. 2017 Jun;28(2):239-242. doi: 10.1007/s00399-017-0513-3. Epub 2017 May 31. Herzschrittmacherther Elektrophysiol. 2017. PMID: 28567490 Review. English.
-
Combined third-generation implantable cardioverter defibrillator with permanent unipolar pacemakers: preliminary observations.Pacing Clin Electrophysiol. 1996 Feb;19(2):136-42. doi: 10.1111/j.1540-8159.1996.tb03304.x. Pacing Clin Electrophysiol. 1996. PMID: 8834682
-
Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrogram width criterion.Europace. 2001 Jan;3(1):80-4. doi: 10.1053/eupc.2000.0145. Europace. 2001. PMID: 11271957
-
Multicenter investigation of an implantable cardioverter-defibrillator algorithm to detect oversensing.Heart Rhythm. 2017 Jul;14(7):1008-1015. doi: 10.1016/j.hrthm.2017.03.023. Epub 2017 Mar 18. Heart Rhythm. 2017. PMID: 28323174
-
Sensing and detection in Medtronic implantable cardioverter defibrillators.Herzschrittmacherther Elektrophysiol. 2016 Sep;27(3):193-212. doi: 10.1007/s00399-016-0450-6. Epub 2016 Sep 8. Herzschrittmacherther Elektrophysiol. 2016. PMID: 27624809 Review. English.
Cited by
-
T-Wave Oversensing with Contemporary Implantable Cardioverter-Defibrillators.J Cardiovasc Dev Dis. 2023 Oct 17;10(10):430. doi: 10.3390/jcdd10100430. J Cardiovasc Dev Dis. 2023. PMID: 37887877 Free PMC article.
-
Inappropriate pacing due to T-wave oversensing.J Thorac Dis. 2020 May;12(5):2983-2985. doi: 10.21037/jtd.2020.02.19. J Thorac Dis. 2020. PMID: 32642212 Free PMC article. No abstract available.
-
T wave oversensing.Indian Pacing Electrophysiol J. 2010 Jun 5;10(6):236-8. Indian Pacing Electrophysiol J. 2010. PMID: 20552058 Free PMC article. No abstract available.
-
Ventricular rate stabilization algorithm of ICD causing dual chamber pacing during ventricular tachycardia.J Interv Card Electrophysiol. 2003 Dec;9(3):397-400. doi: 10.1023/a:1027412014937. J Interv Card Electrophysiol. 2003. PMID: 14618063
-
Ventricular rate stabilization for treatment of recurrent VT.Herzschrittmacherther Elektrophysiol. 2017 Jun;28(2):239-242. doi: 10.1007/s00399-017-0513-3. Epub 2017 May 31. Herzschrittmacherther Elektrophysiol. 2017. PMID: 28567490 Review. English.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical