Mixing collagen with fibrin glue to strengthen the sealing effect for pulmonary air leakage
Abstract
Background: To strengthen the sealing effect of fibrin glue for pulmonary air leakage, atelocollagen was mixed with the glue and the mixing effect was examined.
Methods: A mixture of fibrinogen and thrombin with atelocollagen was used as a test sample. The concentrations of atelocollagen were adjusted to levels of 0%, 0.375%, 0.75%, 1.1%, and 1.5%. We next performed air leakage tests on a plastic cap with pin holes and swine lung and also measured the elasticity and the adhesion strength.
Results: The pressure required to rupture the sealant on a plastic cap with pin holes increased as the concentration of atelocollagen increased, and the bursting pressures were significantly higher in the glue with 0.75%, 1.1%, and 1.5% of atelocollagen than in the glue without atelocollagen (p < 0.01 and p < 0.001). The air leakage pressure on the swine lung was significantly higher in the glue with 0.375%, 0.75%, and 1.1% of atelocollagen than in the glue without atelocollagen (p < 0.05 and p < 0.01), and it was the highest with 0.75%. The elasticity of the glue significantly increased as the concentration of atelocollagen increased (p < 0.001). However, the adhesion strength of the glue significantly decreased as the concentration of atelocollagen increased (p < 0.05 to p < 0.001).
Conclusions: The mixing of atelocollagen with fibrin glue more effectively sealed pulmonary air leakage due to an increased elasticity of the glue while its adhesion strength decreased. The optimal concentration of atelocollagen in the fibrin glue to obtain the best sealing effect was 0.75%.
Similar articles
-
Application of atelocollagen sheet for sellar reconstruction.J Clin Neurosci. 2016 May;27:142-6. doi: 10.1016/j.jocn.2015.08.039. Epub 2016 Jan 5. J Clin Neurosci. 2016. PMID: 26778360
-
The sealing effect of fibrin glue against alveolar air leakage evaluated up to 48 h; comparison between different methods of application.Eur J Cardiothorac Surg. 2005 Jul;28(1):39-42. doi: 10.1016/j.ejcts.2005.02.044. Eur J Cardiothorac Surg. 2005. PMID: 15923121
-
Gelatin-resorcinol-formaldehyde-glutaraldehyde glue for sealing pulmonary air leaks during thoracoscopic operation.Ann Thorac Surg. 1999 Jan;67(1):212-6. doi: 10.1016/s0003-4975(98)01184-9. Ann Thorac Surg. 1999. PMID: 10086552
-
Fibrin sealant: a review of its use in surgery and endoscopy.Drugs. 1999 Nov;58(5):863-86. doi: 10.2165/00003495-199958050-00010. Drugs. 1999. PMID: 10595866 Review.
-
Fibrin glue as a drug delivery system.J Control Release. 2010 Nov 20;148(1):49-55. doi: 10.1016/j.jconrel.2010.06.025. Epub 2010 Jul 15. J Control Release. 2010. PMID: 20637815 Free PMC article. Review.
Cited by
-
Fibrin mechanical properties and their structural origins.Matrix Biol. 2017 Jul;60-61:110-123. doi: 10.1016/j.matbio.2016.08.003. Epub 2016 Aug 20. Matrix Biol. 2017. PMID: 27553509 Free PMC article. Review.
-
Definition of postresectional residual pleural space.Can J Surg. 2007 Feb;50(1):39-42. Can J Surg. 2007. PMID: 17391615 Free PMC article.
-
Compression-induced structural and mechanical changes of fibrin-collagen composites.Matrix Biol. 2017 Jul;60-61:141-156. doi: 10.1016/j.matbio.2016.10.007. Epub 2016 Oct 15. Matrix Biol. 2017. PMID: 27751946 Free PMC article.
-
Type I Collagen-Fibrin Mixed Hydrogels: Preparation, Properties and Biomedical Applications.Gels. 2020 Oct 20;6(4):36. doi: 10.3390/gels6040036. Gels. 2020. PMID: 33092154 Free PMC article. Review.
-
Surgical sealant for preventing air leaks after pulmonary resections in patients with lung cancer.Cochrane Database Syst Rev. 2010 Jan 20;2010(1):CD003051. doi: 10.1002/14651858.CD003051.pub3. Cochrane Database Syst Rev. 2010. PMID: 20091536 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous