Analysis of magnetic resonance imaging acoustic noise generated by a 4.7 T experimental system
- PMID: 11099151
- DOI: 10.1080/000164800750000270
Analysis of magnetic resonance imaging acoustic noise generated by a 4.7 T experimental system
Abstract
High intensity acoustic noise is an undesirable side-effect in magnetic resonance imaging (MRI) that can cause discomfort and hearing loss in patients and may be an impediment in functional MRI (fMRI) studies of the auditory system. Experimental MRI systems with high magnetic field strengths may generate acoustic noise of higher sound pressure levels (SPLs) than conventional 1.0 and 1.5 T clinical systems. We measured the SPL and spectral content of the acoustic noise generated by the Bruker Biospect 47/40 4.7 T experimental MRI system during scanning sequences commonly used in animal testing. Each sequence generated acoustic noise of high SPL, rapid pulse rates, amplitude-modulated pulse envelopes and multi-peaked spectra. The rapid acquisition with enhancement sequence with a 0.25 mm slice thickness generated SPLs of up to 129 dB peak SPL and 130 dB (A). Fourier analysis of the spectral content of the acoustic noise generated by each MRI sequence showed a wide band of acoustic energy with spectral peaks from 0.2-5 kHz. The intense MRI acoustic impulse noise generated by the 4.7 T system may cause masking of stimuli used in fMRI of the auditory cortex, reduce the hearing acuity of experimental animals and present a risk for unprotected human ears.
Similar articles
-
MRI acoustic noise: sound pressure and frequency analysis.J Magn Reson Imaging. 1997 May-Jun;7(3):606-11. doi: 10.1002/jmri.1880070327. J Magn Reson Imaging. 1997. PMID: 9170052
-
Acoustic noise generated by TMS in typical environment and inside an MRI scanner.Brain Stimul. 2024 Mar-Apr;17(2):184-193. doi: 10.1016/j.brs.2024.02.006. Epub 2024 Feb 9. Brain Stimul. 2024. PMID: 38342363
-
Acoustic noise during functional magnetic resonance imaging.J Acoust Soc Am. 2000 Oct;108(4):1683-96. doi: 10.1121/1.1310190. J Acoust Soc Am. 2000. PMID: 11051496 Free PMC article.
-
Acoustic noise concerns in functional magnetic resonance imaging.Hum Brain Mapp. 2003 Nov;20(3):123-41. doi: 10.1002/hbm.10134. Hum Brain Mapp. 2003. PMID: 14601139 Free PMC article. Review.
-
Methodological challenges and solutions in auditory functional magnetic resonance imaging.Front Neurosci. 2014 Aug 21;8:253. doi: 10.3389/fnins.2014.00253. eCollection 2014. Front Neurosci. 2014. PMID: 25191218 Free PMC article. Review.
Cited by
-
Model gradient coil employing active acoustic control for MRI.MAGMA. 2007 Dec;20(5-6):223-31. doi: 10.1007/s10334-007-0086-y. Epub 2007 Nov 15. MAGMA. 2007. PMID: 18004601 Free PMC article.
-
Using neuroimaging to understand the cortical mechanisms of auditory selective attention.Hear Res. 2014 Jan;307:111-20. doi: 10.1016/j.heares.2013.06.010. Epub 2013 Jul 9. Hear Res. 2014. PMID: 23850664 Free PMC article. Review.
-
Acoustic-noise-optimized diffusion-weighted imaging.MAGMA. 2015 Dec;28(6):511-21. doi: 10.1007/s10334-015-0492-5. Epub 2015 Jun 20. MAGMA. 2015. PMID: 26092411
-
Assessing the influence of scanner background noise on auditory processing. I. An fMRI study comparing three experimental designs with varying degrees of scanner noise.Hum Brain Mapp. 2007 Aug;28(8):703-20. doi: 10.1002/hbm.20298. Hum Brain Mapp. 2007. PMID: 17080440 Free PMC article.
-
Assessing the influence of scanner background noise on auditory processing. II. An fMRI study comparing auditory processing in the absence and presence of recorded scanner noise using a sparse design.Hum Brain Mapp. 2007 Aug;28(8):721-32. doi: 10.1002/hbm.20299. Hum Brain Mapp. 2007. PMID: 17089376 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous