Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers

Affiliations

Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers

S Shimizu et al. Int J Radiat Oncol Biol Phys. .

Abstract

Purpose: A real-time tracking radiotherapy was investigated to assess its usefulness in precise localization and verification of prostate and bladder cancers.

Methods and materials: The real-time tracking radiation therapy (RTRT) system consists of implantation of a 2.0-mm gold marker into a clinical target volume (CTV), three-dimensional radiation treatment planning (3DRTP) system, and the use of two sets of diagnostic x-ray television systems in the linear accelerator room, image processing units, and an image display unit. The position of the patient can be corrected by adjusting the actual marker position to the planned marker position, which has been transferred from the 3DRTP and superimposed on the fluoroscopic image on the display unit of the RTRT system. The position of the markers can be visualized during irradiation and after treatment delivery to verify the accuracy of the localization. Ten patients with prostate cancer and 5 patients with bladder cancer were examined using this system for the treatment setup on 91 occasions.

Results: After manual setup using skin markers, the median of absolute value of discrepancies between the actual position of the marker and the planned position of the marker for prostate cancer was 3.4 (0.1-8.9) mm, 4.1 (0.2-18.1) mm, and 2.3 (0.0-10.6) mm for the lateral, anteroposterior, and craniocaudal directions, respectively. The 3D median distance between the actual and planned positions of the marker was 6.9 (1.1-18.2) mm for prostate cancer and 6.9 (1.7-18.6) mm for bladder cancer. After relocation using RTRT, the 3D distance between the actual and planned position of the marker was 0.9 +/- 0.9 mm. Median 3D distances between actual positions after treatment delivery and planned positions were 1.6 (0.0-6.3) mm and 2.0 (0.5-8.0) mm during daily radiotherapy for the marker in patients with prostate cancer and bladder cancer, respectively.

Conclusion: We believe the new positioning system can reduce uncertainty due to setup error and internal organ motion, although further improvement is needed for the system to account for the rotational and elastic changes of the affected tissues.