What is the optimum leaf width of a multileaf collimator?
- PMID: 11128300
- DOI: 10.1118/1.1319524
What is the optimum leaf width of a multileaf collimator?
Abstract
The following question is investigated: How narrow do the leaves of a multileaf collimator have to be such that further reduction of the leaf width does not lead to physical improvements of the dose distribution. Because of the physical principles of interaction between radiation and matter, dose distributions in radiotherapy are generally relatively smooth. According to the theory of sampling, the dose distribution can therefore be represented by a set of evenly spaced samples. The distance between the samples is identified with the distance between the leaf centers of a multileaf collimator. The optimum sampling distance is derived from the 20% to 80% field edge penumbra through the concept of the dose deposition kernel, which is approximated by a Gaussian. The leaf width of the multileaf collimator is considered to be independent from the sampling distance. Two cases are studied in detail: (i) the leaf width equals the sampling distance, which is the regular case, and (ii) the leaf width is twice the sampling distance. The practical delivery of the latter treatment geometry requires a couch movement or a collimator rotation. The optimum sampling distance equals the 20%-80% penumbra divided by 1.7 and is on the order of 1.5-2 mm for a typical 6 MV beam in soft tissue. The optimum leaf width equals this sampling distance in the regular case. A relatively small deterioration results if the leaf width is doubled, while the sampling distance remains the same. The deterioration can be corrected for by deconvolving the fluence profile with an inverse filter.
Conclusions: With the help of the sampling theory and, more generally, the theory of linear systems, one can find a general answer to the question about the optimum leaf width of a multileaf collimator from a physical point of view. It is important to distinguish between the sampling distance and the leaf width. The sampling distance is more critical than the leaf width. The leaf width can be up to twice as large as the sampling width. Furthermore, the derived sampling distance can be used to select the optimum resolution of both the fluence and the dose grid in dose calculation and inverse planning algorithms.
Similar articles
-
Improving the resolution of dynamic intensity modulated radiation therapy delivery by reducing the multileaf collimator sampling distance.Med Phys. 2003 Oct;30(10):2793-803. doi: 10.1118/1.1609992. Med Phys. 2003. PMID: 14596316
-
Effect of multileaf collimator leaf width on physical dose distributions in the treatment of CNS and head and neck neoplasms with intensity modulated radiation therapy.Med Phys. 2002 Jun;29(6):1116-9. doi: 10.1118/1.1481515. Med Phys. 2002. PMID: 12094981
-
A topographic leaf-sequencing algorithm for delivering intensity modulated radiation therapy.Med Phys. 2006 Aug;33(8):2751-6. doi: 10.1118/1.2216876. Med Phys. 2006. PMID: 16964850
-
The effect of leaf width and sampling distance on the "stair-stepping" of field borders defined by multileaf collimators.Australas Phys Eng Sci Med. 2003 Jun;26(2):45-52. Australas Phys Eng Sci Med. 2003. PMID: 12956185
-
Design of a fast multileaf collimator for radiobiological optimized IMRT with scanned beams of photons, electrons, and light ions.Med Phys. 2007 Mar;34(3):877-88. doi: 10.1118/1.2431082. Med Phys. 2007. PMID: 17441233
Cited by
-
Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets.J Med Phys. 2011 Apr;36(2):72-7. doi: 10.4103/0971-6203.79690. J Med Phys. 2011. PMID: 21731222 Free PMC article.
-
Finer leaf resolution and steeper beam edges using a virtual isocentre in concurrence to PTV-shaped collimators in standard distance - a planning study.Radiat Oncol. 2017 May 25;12(1):88. doi: 10.1186/s13014-017-0826-8. Radiat Oncol. 2017. PMID: 28545556 Free PMC article.
-
Dosimetric performance of the new high-definition multileaf collimator for intracranial stereotactic radiosurgery.J Appl Clin Med Phys. 2010 Jun 21;11(3):3040. doi: 10.1120/jacmp.v11i3.3040. J Appl Clin Med Phys. 2010. PMID: 20717077 Free PMC article. Clinical Trial.
-
Implications of a high-definition multileaf collimator (HD-MLC) on treatment planning techniques for stereotactic body radiation therapy (SBRT): a planning study.Radiat Oncol. 2009 Jul 10;4:22. doi: 10.1186/1748-717X-4-22. Radiat Oncol. 2009. PMID: 19591687 Free PMC article.
-
Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.J Appl Clin Med Phys. 2012 May 10;13(3):3729. doi: 10.1120/jacmp.v13i3.3729. J Appl Clin Med Phys. 2012. PMID: 22584170 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources