. 2001 Feb;119(2):562-4.
doi: 10.1378/chest.119.2.562.

Home ventilator low-pressure alarms fail to detect accidental decannulation with pediatric tracheostomy tubes

Affiliations

Home ventilator low-pressure alarms fail to detect accidental decannulation with pediatric tracheostomy tubes

S S Kun et al. Chest. 2001 Feb.

Abstract

Background: Positive-pressure ventilators are equipped with low-inspiratory-pressure alarms to protect patients from hypoventilation. Small uncuffed tracheostomy tubes have a high resistance, and may not trigger these alarms during decannulation.

Study objective: To determine whether ventilator low-inspiratory-pressure alarms are effective in detecting tracheostomy decannulation.

Design: We connected tracheostomy tubes of varying inner diameters (3.0 to 6.0 mm) to a home ventilator and simulated decannulation using low (tidal volume [VT], 600 mL; peak inspiratory pressure [PIP], 25 cm H(2)O), medium (VT, 800 mL; PIP, 30 cm H(2)O), and high (VT, 1,000 mL; PIP, 35 cm H(2)O) ventilator settings.

Results: When the ventilator low-inspiratory-pressure alarm was set at 4 cm H(2)O below the desired PIP, it failed to alarm for simulated decannulation of tracheostomy tubes < 4.5 mm on low and medium settings, and < 4.0 mm on high settings. When the ventilator low-inspiratory-pressure alarm was set at 10 cm H(2)O below the desired PIP, it failed to alarm with tracheostomy tubes < 6.0 mm.

Conclusion: We conclude that ventilator low-inspiratory-pressure alarms fail to alarm during simulated decannulation with small tracheostomy tubes commonly used in children. We speculate that low-inspiratory-pressure alarms set at 4 cm H(2)O below the desired PIP will detect more decannulation than when set at 10 cm H(2)O below the desired PIP.