Correlation between high-pass resolution perimetry and standard threshold perimetry in subjects with glaucoma and ocular hypertension
- PMID: 11196128
- DOI: 10.1023/a:1026573909370
Correlation between high-pass resolution perimetry and standard threshold perimetry in subjects with glaucoma and ocular hypertension
Abstract
Purpose: To evaluate the correlation between High-Pass Resolution Perimetry (HRP) and standard threshold perimetry in patients with glaucoma or ocular hypertension.
Methods: 31 glaucomatous patients and 37 ocular hypertension subjects with previous perimetric examination experience were consecutively recruited and only one eye for each patient was selected at random. Glaucomatous patients were classified as having primary open angle glaucoma when they had an abnormal visual field and/or an abnormal optic nerve head (ONH)/retinal nerve fiber layer (RNFL) typical of glaucoma, open angle at gonioscopy and no clinically apparent secondary cause for their glaucoma. Ocular hypertension subjects were defined as having intraocular pressure >21 mm Hg on no treatment, normal visual field, normal ONH and RNFL, elevated intraocular pressure without any treatment. All the subjects were examined with Humphrey Field Analyzer (HFA) 640, 'program central 30-2' (Humphrey Systems, San Leandro, CA, USA) and with High-Pass Resolution Perimeter (HRP), Ophthimus version 2.4,'ring program' (Nikon-HighTech Vision, Goteborg, Sweden). Visual field indices were obtained with both systems: for HFA mean deviation (MD), corrected pattern standard deviation (CPSD) and short term fluctuation (SF), while for HRP global deviation (GD), local deviation (LD), form index (FI) and neural capacity (NC). The data were analyzed by descriptive analysis, Student's t test with Bonferroni's correction or Mann-Whitney non-parametric test and Pearson or Spearman's correlation coefficient.
Results: A significant correlation was found between MD and GD (r = -0.81), CPSD and LD (r = 0.87), PSD and LD (r = 0.72). NC was significantly correlated with MD (r = 0.76), GD (r = -0.94). FI was significantly correlated with PSD (r = -0.58), CPSD (r = -0.72), LD (r = -0.56). When the same data were analyzed for the glaucomatous group only, similar results were found; in the ocular hypertensive group no significant correlation was found except between NC and MD (r = 0.52).
Conclusion: HRP indices vary comparably with HFA indices. Parameters as NC and FI were significantly correlated with standard visual field indices of both HFA and HRP. Although the clinical applications for FI are not clear yet, NC could detect both early glaucomatous damage and age related changes.
Similar articles
-
Comparison between relative dispersion analysis of high-pass resolution perimetry and standard threshold perimetry.Eye (Lond). 2000 Oct;14 Pt 5:742-6. doi: 10.1038/eye.2000.195. Eye (Lond). 2000. PMID: 11116696
-
Detection of glaucomatous visual field defect by nonconventional perimetry.Am J Ophthalmol. 2003 Jan;135(1):35-9. doi: 10.1016/s0002-9394(02)01818-4. Am J Ophthalmol. 2003. PMID: 12504694 Clinical Trial.
-
Blue-on-yellow visual field and retinal nerve fiber layer in ocular hypertension and glaucoma.Ophthalmology. 1998 Nov;105(11):2077-81. doi: 10.1016/S0161-6420(98)91128-9. Ophthalmology. 1998. PMID: 9818609 Clinical Trial.
-
The value of high-pass resolution perimetry in glaucoma.Curr Opin Ophthalmol. 2000 Apr;11(2):85-9. doi: 10.1097/00055735-200004000-00003. Curr Opin Ophthalmol. 2000. PMID: 10848225 Review.
-
Diagnostic value of short-wavelength automated perimetry.Curr Opin Ophthalmol. 1996 Apr;7(2):54-8. doi: 10.1097/00055735-199604000-00010. Curr Opin Ophthalmol. 1996. PMID: 10163323 Review.
Cited by
-
Simplified automatic method for measuring the visual field using the perimeter ZERK 1.Biomed Eng Online. 2016 Jul 25;15(1):88. doi: 10.1186/s12938-016-0210-1. Biomed Eng Online. 2016. PMID: 27456974 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous