Defects in rotary nickel-titanium files after clinical use
- PMID: 11199711
- DOI: 10.1097/00004770-200003000-00008
Defects in rotary nickel-titanium files after clinical use
Abstract
The purpose of this study was to analyze the type and frequency of defects in nickel-titanium rotary endodontic files after routine clinical use, and to draw conclusions regarding the reasons for failure. All of the files (total: 378, Quantec Series 2000) discarded after normal use from a specialist endodontic practice over 6 months were analyzed. Almost 50% of the files showed some visible defect; 21% were fractured and 28% showed other defects without fracture. Fractured files could be divided into two groups according to the characteristics of the defects observed. Torsional fracture occurred in 55.7% of all fractured files, whereas flexural fatigue occurred in 44.3%. The results indicated that torsional failure, which may be caused by using too much apical force during instrumentation, occurred more frequently than flexural fatigue, which may result from use in curved canals.
Similar articles
-
Deformation and fracture of Mtwo rotary nickel-titanium instruments after clinical use.J Endod. 2009 Oct;35(10):1396-9. doi: 10.1016/j.joen.2009.06.014. Epub 2009 Aug 15. J Endod. 2009. PMID: 19801238
-
Influence of electrochemical polishing on the mechanical properties of K3 nickel-titanium rotary instruments.J Endod. 2008 Dec;34(12):1533-6. doi: 10.1016/j.joen.2008.08.023. Epub 2008 Sep 26. J Endod. 2008. PMID: 19026889
-
Defects in ProTaper S1 instruments after clinical use: fractographic examination.Int Endod J. 2005 Nov;38(11):802-9. doi: 10.1111/j.1365-2591.2005.01020.x. Int Endod J. 2005. PMID: 16218972
-
A review of cyclic fatigue testing of nickel-titanium rotary instruments.J Endod. 2009 Nov;35(11):1469-76. doi: 10.1016/j.joen.2009.06.015. Epub 2009 Sep 1. J Endod. 2009. PMID: 19840633 Review.
-
Understanding and evaluating the endodontic file.Gen Dent. 2000 Nov-Dec;48(6):690-2. Gen Dent. 2000. PMID: 12004664 Review.
Cited by
-
Fracture Incidence of New Reciprocating Nickel-Titanium (NiTi) Files: A Cross-Sectional Retrospective Study.Cureus. 2024 Aug 25;16(8):e67762. doi: 10.7759/cureus.67762. eCollection 2024 Aug. Cureus. 2024. PMID: 39323712 Free PMC article.
-
In Vivo Evaluation of Operative Torque Generated by Two Nickel-Titanium Rotary Instruments during Root Canal Preparation.Eur J Dent. 2019 Oct;13(4):556-562. doi: 10.1055/s-0039-1698369. Epub 2019 Dec 31. Eur J Dent. 2019. PMID: 31891973 Free PMC article.
-
Analysis of the Importance of the Motion Used in the Resistance of Different Mechanical Instrumentation Systems in Endodontics: A Comparative Study.Materials (Basel). 2022 Jun 24;15(13):4443. doi: 10.3390/ma15134443. Materials (Basel). 2022. PMID: 35806567 Free PMC article.
-
In vitro comparison of cyclic fatigue resistance of TruNatomy in single and double curvature canals compared with different nickel-titanium rotary instruments.BMC Oral Health. 2020 Feb 4;20(1):38. doi: 10.1186/s12903-020-1027-7. BMC Oral Health. 2020. PMID: 32019522 Free PMC article.
-
Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments.Clin Oral Investig. 2018 Sep;22(7):2633-2638. doi: 10.1007/s00784-018-2362-9. Epub 2018 Feb 1. Clin Oral Investig. 2018. PMID: 29392414
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources