Radiologist detection of microcalcifications with and without computer-aided detection: a comparative study
- PMID: 11222075
- DOI: 10.1053/crad.2000.0592
Radiologist detection of microcalcifications with and without computer-aided detection: a comparative study
Abstract
Aim: To compare the sensitivity and specificity of microcalcification detection by radiologists alone and assisted by a computer-aided detection (CAD) system.
Materials and methods: Films of 106 patients were masked, randomized, digitized and analysed by the CAD-system. Five readers interpreted the original mammograms and were blinded to demographics, medical history and earlier films. Forty-two mammograms with malignant microcalcifications, 40 with benign microcalcifications and 24 normal mammograms were included. Results were recorded on a standardized image interpretation form. The mammograms with suspicious areas flagged by the CAD-system were displayed on mini-monitors and immediately re-reviewed. The interpretation was again recorded on a new copy of the standard form and classified according to six groups.
Results: Forty-one out of 42 (98%) malignant microcalcifications and 32 of 40 (80%) benign microcalcifications were flagged by the CAD-system. There was an average of 1.2 markers per image. The sensitivity for malignant microcalcifications detection by mammographers without and with the CAD-system ranged from 81% to 98% and from 88% to 98%, respectively. The mean difference without and with CAD-system was 2.2% (range 0-7%).
Conclusion: No statistically significant changes in sensitivity were found when experienced mammographers were assisted by the CAD-system, with no significant compromise in specificity.
Copyright 2001 The Royal College of Radiologists.
Similar articles
-
Evaluation of a wavelet-based computer-assisted detection system for identifying microcalcifications in digital full-field mammography.Acta Radiol. 2004 Apr;45(2):136-41. doi: 10.1080/02841850410003770. Acta Radiol. 2004. PMID: 15191095
-
[Application of a computer-aided detection (CAD) system to digitalized mammograms for identifying microcalcifications].Radiol Med. 2001 May;101(5):334-40. Radiol Med. 2001. PMID: 11438784 Italian.
-
Comparison of independent double readings and computer-aided diagnosis (CAD) for the diagnosis of breast calcifications.Acad Radiol. 2006 Jan;13(1):84-94. doi: 10.1016/j.acra.2005.09.086. Acad Radiol. 2006. PMID: 16399036
-
[Computer-aided diagnosis of breast cancer].Nihon Igaku Hoshasen Gakkai Zasshi. 2002 Jul;62(8):409-14. Nihon Igaku Hoshasen Gakkai Zasshi. 2002. PMID: 12187836 Review. Japanese.
-
[Understanding CAD (computer-aided diagnosis) in mammography].J Radiol. 2005 Jan;86(1):29-35. doi: 10.1016/s0221-0363(05)81319-8. J Radiol. 2005. PMID: 15785414 Review. French.
Cited by
-
Breast imaging: how we manage diagnostic technology at a multidisciplinary breast center.J Oncol. 2012;2012:213421. doi: 10.1155/2012/213421. Epub 2012 Jul 5. J Oncol. 2012. PMID: 22829820 Free PMC article.
-
Risk-Aware Machine Learning Classifier for Skin Lesion Diagnosis.J Clin Med. 2019 Aug 17;8(8):1241. doi: 10.3390/jcm8081241. J Clin Med. 2019. PMID: 31426482 Free PMC article.
-
CAD for mammography: the technique, results, current role and further developments.Eur Radiol. 2006 Jul;16(7):1449-60. doi: 10.1007/s00330-005-0089-x. Epub 2006 Jan 17. Eur Radiol. 2006. PMID: 16416275 Review.
-
Analog Computer-Aided Detection (CAD) information can be more effective than binary marks.Atten Percept Psychophys. 2017 Feb;79(2):679-690. doi: 10.3758/s13414-016-1250-0. Atten Percept Psychophys. 2017. PMID: 27928658 Free PMC article.
-
Influence of computer-aided detection on performance of screening mammography.N Engl J Med. 2007 Apr 5;356(14):1399-409. doi: 10.1056/NEJMoa066099. N Engl J Med. 2007. PMID: 17409321 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous