Improved multi-station peripheral MR angiography with a dedicated vascular coil
- PMID: 11241825
- DOI: 10.1002/jmri.1069
Improved multi-station peripheral MR angiography with a dedicated vascular coil
Abstract
Delineation of small branch vessels can be crucial for assessing the peripheral arterial system of patients requiring surgical grafting. Thus signal-to-noise needs to be maximized. We evaluated the performance of a dedicated peripheral vascular coil in four subjects by comparing it to the body coil using DSA as the standard of reference. SNR and CNR values of the dedicated peripheral coil exceeded those obtained with the body coil by a mean of 398%, thus permitting improved delineation of the infrapopliteal arterial morphology.