Comparative Study
. 2001 Mar;52(3):322-6; discussion 327.
doi: 10.1002/ccd.1074.

Comparison of myocardial fractional flow reserve and intravascular ultrasound for the assessment of slotted-tube stents

Affiliations
Comparative Study

Comparison of myocardial fractional flow reserve and intravascular ultrasound for the assessment of slotted-tube stents

D G Katritsis et al. Catheter Cardiovasc Interv. 2001 Mar.

Abstract

Intravascular ultrasound (IVUS) and myocardial fractional flow reserve (FFR) have been reported to provide similar results for assessment of coil stent deployment. Their relative value in slotted-tube stents has not been investigated. Fourteen patients subjected to coronary angioplasty and IVUS-guided elective stenting with a slotted-tube stent underwent IVUS assessment and FFR measurement following stent implantation at inflation pressures of 12 and 18 atm. FFR values (mean +/- SD) preangioplasty, postangioplasty, and poststenting at 12 atm and 18 atm, were 0.58 +/- 0.07, 0.83 +/- 0.05, 0.94 +/- 0.02, and 0.94 +/- 0.02, respectively. After inflation at 12 atm, the area under the receiver operating characteristic (ROC) curve for the concordance of IVUS and FFR measurements was 0.89 (P = 0.02). Six patients had either an abnormal IVUS (n = 2) or FFR < 0.94 (n = 1) or both abnormal IVUS and FFR < 0.94 (n = 3) after the first inflation and had a second inflation at 18 atm. The area under the ROC curve for the concordance between IVUS and FFR final measurements was 0.855 (P = 0.10). Perfect concordance between IVUS and FFR was seen only for FFR values less than 0.91 or larger than 0.94. Overall, IVUS and FFR have substantial concordance with respect to slotted-tube stent deployment. However, FFR values between 0.91 and 0.94 after inflation are difficult to interpret.