Engineering considerations in the design of an ankle joint
- PMID: 1125359
Engineering considerations in the design of an ankle joint
Abstract
A prothesis has been designed to replace the articulating surfaces of the human ankle joint. The prothesis is in two parts, each forming a segment of a right circular cylinder with a single axis of rotation. The concave tibial component is manufactured from ultra-high molecular weight polyethylene and the talar component is manufactured from medical grade stainless steel. It is likely, however, that the talar component will be commercially manufactured from cobalt chrome alloy (Vitallium or Vinertia). The two components are secured to the cancellous bone by polymethylmethacrylate bone cement and laboratory tests have indicated that the bond should be strong enough to withstand the loads encountered at the ankle joint in vivo. The tests have also shown that the stability and strength of the ankle are not seriously reduced by implantation of the prosthesis. Laboratory wear tests and clinical experience over the last two years encourage optimism over the long term performance of the prothesis.
Similar articles
-
Survivorship and clinical evaluation of cementless, meniscal-bearing total ankle replacements.Semin Arthroplasty. 1992 Jan;3(1):43-50. Semin Arthroplasty. 1992. PMID: 10147571
-
Ankle arthroplasty. A comparative study of cemented metal and uncemented ceramic prostheses.Clin Orthop Relat Res. 1990 Mar;(252):209-16. Clin Orthop Relat Res. 1990. PMID: 2302887
-
[Wear characteristics of different metal-polyethylene beating surfaces. An experimental study of a new model of knee prosthesis].Rev Chir Orthop Reparatrice Appar Mot. 1996;82(6):522-8. Rev Chir Orthop Reparatrice Appar Mot. 1996. PMID: 9122523 French.
-
Particulate metallic debris in cemented total hip arthroplasty.Clin Orthop Relat Res. 1993 Aug;(293):160-73. Clin Orthop Relat Res. 1993. PMID: 8339477 Review.
-
Characteristics of metal and ceramic total hip bearing surfaces and their effect on long-term ultra high molecular weight polyethylene wear.Clin Orthop Relat Res. 1993 Sep;(294):361-78. Clin Orthop Relat Res. 1993. PMID: 8358943 Review.
Cited by
-
Biomechanics of the natural, arthritic, and replaced human ankle joint.J Foot Ankle Res. 2014 Feb 6;7(1):8. doi: 10.1186/1757-1146-7-8. J Foot Ankle Res. 2014. PMID: 24499639 Free PMC article.
-
Ultra-high-molecular-weight Polyethylene (UHMWPE) Wing Method for Strong Cranioplasty.Neurol Med Chir (Tokyo). 2021 Sep 15;61(9):549-556. doi: 10.2176/nmc.oa.2021-0032. Epub 2021 Jun 7. Neurol Med Chir (Tokyo). 2021. PMID: 34092749 Free PMC article.
-
Dynamic simulation of the natural and replaced human ankle joint.Med Biol Eng Comput. 2002 Mar;40(2):193-9. doi: 10.1007/BF02348124. Med Biol Eng Comput. 2002. PMID: 12043800
-
The production and biology of polyethylene wear debris.Arch Orthop Trauma Surg (1978). 1978 May 30;91(3):167-81. doi: 10.1007/BF00379748. Arch Orthop Trauma Surg (1978). 1978. PMID: 666546
-
Effects of positioning on radiographic measurements of ankle morphology: a computerized tomography-based simulation study.Biomed Eng Online. 2013 Dec 20;12:131. doi: 10.1186/1475-925X-12-131. Biomed Eng Online. 2013. PMID: 24359413 Free PMC article.