Autoadjusting CPAP therapy based on impedance efficacy, compliance and acceptance
- PMID: 11254519
- DOI: 10.1164/ajrccm.163.3.2006168
Autoadjusting CPAP therapy based on impedance efficacy, compliance and acceptance
Abstract
Constant continuous positive airway pressure (CPAP) is the treatment of choice for the obstructive sleep apnea syndrome (OSAS). To enable the pressure to be matched more accurately to actual requirements, and thus increase patient acceptance, an autoadjusting device based on the measurement of upper airway impedance was developed (APAP(FOT)). We investigated the efficacy and compliance in continuous use at home. Fifty-two patients were treated (randomized crossover) with CPAP and APAP(FOT) for 6 wk each. Respiratory disturbances, sleep profile, and arousals improved significantly with both modes (AHI: baseline, 35.1 +/- 26/h; APAP(FOT), 5.0 +/- 5.2; CPAP, 4.3 +/- 6.3; p < 0.001 baseline versus each mode). The mean pressure with APAP(FOT) was significantly reduced as compared with CPAP (CPAP, 7.8 +/- 1.5 cm H2O; APAP(FOT), 5.7 +/- 1.8 cm H2O; p < 0.001). Under APAP(FOT) the pressure was lower than that under CPAP for 81.5 +/- 21% of the time. Although overall use did not differ, 75% of the patients preferred APAP(FOT) for home treatment. We conclude that APAP(FOT) is as efficacious as constant CPAP in the treatment of OSAS. The treatment pressure can be reduced significantly, and sleep microstructure improved with APAP(FOT). These might be the reasons for patient preference of automatic therapy.
Comment in
-
Autoadjusting continuous positive airway pressure: what can we expect?Am J Respir Crit Care Med. 2001 May;163(6):1295-6. doi: 10.1164/ajrccm.163.6.ed1801f. Am J Respir Crit Care Med. 2001. PMID: 11371386 Review. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical