Comparative Study
. 2001 Mar;21(2):101-7.

Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in adults

Affiliations
  • PMID: 11261343
Comparative Study

Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in adults

E A Mallen et al. Ophthalmic Physiol Opt. 2001 Mar.

Abstract

A clinical evaluation of the Shin-Nippon SRW-5000 (Japan), a newly released commercial autorefractor, was undertaken to assess its repeatability and validity compared to subjective refraction. Measurements of refractive error were performed on 200 eyes of 100 subjects (aged 24.4 +/- 8.0 years) subjectively (non-cycloplegic) by one optometrist and objectively with the SRW-5000 autorefractor by a second. Repeatability was assessed by examining the differences between the seven autorefractor readings taken from each eye and by re-measuring the objective prescription of 50 eyes at a subsequent session. Although the SRW-5000 read slightly more plus than subjective refraction (mean spherical equivalent +0.16 +/- 0.44 D), it was found to be highly valid (accurate) compared to subjective refraction and repeatable over the prescription range of +6.50 to -15.00 D examined. The Shin-Nippon SRW-5000 autorefractor is therefore a valuable complement to subjective refraction and as it offers the advantage of a binocular open field-of-view, has a great potential benefit for accommodation research studies.

PubMed Disclaimer