Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1975 May;38(5):919-21.
doi: 10.1152/jappl.1975.38.5.919.

Deep body temperature from aural canal with servo-controlled heating to outer ear

Comparative Study

Deep body temperature from aural canal with servo-controlled heating to outer ear

W R Keatinge et al. J Appl Physiol. 1975 May.

Abstract

A portable battery-powered device was used to measure temperature in the external aural canal by a thermistor while keeping the temperature of the outer ear, monitored by a second thermistor at the same level by servo-controlled heating. Aural canal temperature then always stabilized within 0.35 degrees C of esophageal in moving air at between 18 and 45 degrees C. It often did so in colder air, but stabilization was slow even after brief exposure to the cold air. Aural temperature then paralleled esophageal within 0.35 degrees C during rapid changes in deep body temperature while rectal temperature was seriously depressed in cold air. These conclusions were based on 30 experiments on 11 male and 7 female subjects.

PubMed Disclaimer

Publication types

LinkOut - more resources