Comparative Study
. 2000 Dec;26(12):1837-42.
doi: 10.1007/s001340000705.

Accuracy of totally implanted ports, tunnelled, single- and multiple-lumen central venous catheters for measurement of central venous pressure

Affiliations
Comparative Study

Accuracy of totally implanted ports, tunnelled, single- and multiple-lumen central venous catheters for measurement of central venous pressure

F Blot et al. Intensive Care Med. 2000 Dec.

Erratum in

  • Intensive Care Med 2001 May;27(5):953

Abstract

Objective: To verify the accuracy of totally implanted ports, tunnelled central venous catheters (CVC), widely used in cancer patients, and multi-lumen catheters, used in intensive care units (ICUs), in measuring central venous pressure (CVP), using right atrial pressure (RAP) measured in a Swan-Ganz catheter as the reference standard.

Design: A prospective study, over a 10-month period.

Setting: A medical-surgical ICU in a comprehensive cancer centre.

Patients and participants: Patients who had both (1) a Swan-Ganz catheter and (2) either a tunnelled catheter, a single or a multi-lumen catheter, or a totally implanted port.

Interventions: RAP and CVP were measured simultaneously in each patient.

Measurements and results: Fifty-six pairs of RAP-CVP measurements were performed in 35 patients: 6 tunnelled catheters, 6 non-tunnelled single-lumen catheters, 26 multiple-lumen catheters and 18 totally implanted ports were studied. RAP measured in the Swan-Ganz catheter and CVP measured in the CVC were strongly correlated (r = 0.94, p < 0.01), whatever the type of catheter studied. The mean difference between RAP and CVP was -0.39 +/- 1.73 (SD) mmHg. In 51 cases (91%), the difference was within the limits of agreement (-3.78 to 3.00 mmHg, Bland and Altman method). For the five cases with a difference of 4 mmHg (three totally implanted ports, one double- and one triple-lumen catheter), CVP was greater than RAP.

Conclusions: CVP can be accurately measured in totally implanted ports, tunnelled or non-tunnelled single-lumen and multiple-lumen catheters. When the difference exceeds the limit of agreement, the discrepancy between the two measurements has limited significance in most cases.