Highly sensitive, automated immunoassay for immunoglobulin free light chains in serum and urine
- PMID: 11274017
Highly sensitive, automated immunoassay for immunoglobulin free light chains in serum and urine
Abstract
Background: Bence Jones proteins or monoclonal immunoglobulin kappa and lambda free light chains (FLCs) are important markers for identifying and monitoring many patients with B-cell tumors. Automated immunoassays that measure FLCs in urine and serum have considerable clinical potential.
Methods: Sheep antibodies, specific for FLCs, were prepared by immunization with pure kappa and lambda molecules and then adsorbed extensively against whole immunoglobulins. The antibodies were conjugated onto latex particles and used to assay kappa and lambda FLCs on the Beckman IMMAGE protein analyzer.
Results: The unconjugated antibodies showed minimal cross-reactivity with intact immunoglobulins or other proteins. With latex-conjugated antibodies, kappa and lambda FLCs could be measured in normal sera and most normal urine samples. Patients with multiple myeloma had increased concentrations of the relevant serum FLC, whereas both FLCs were increased in the sera of patients with systemic lupus erythematosus.
Conclusions: We developed sensitive, automated immunoassays for kappa and lambda FLC measurements in serum and urine that should facilitate the assessment of patients with light chain abnormalities.
Comment in
-
Measurement of free light chains in urine.Clin Chem. 2001 Nov;47(11):2069-70. Clin Chem. 2001. PMID: 11673389 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources