High-resolution blood flow velocity measurements in the human finger
- PMID: 11284002
- DOI: 10.1002/mrm.1096
High-resolution blood flow velocity measurements in the human finger
Abstract
MR phase contrast blood flow velocity measurements in the human index finger were performed with triggered, nontriggered, and cine acquisition schemes. A strong (G(max) = 200 mT/m), small bore (inner diameter 12 cm) gradient system inserted in a whole body 3 Tesla MR scanner allowed high-resolution imaging at short echo times, which decreases partial volume effects and flow artifacts. Arterial blood flow velocities ranging from 4.9-19 cm/sec were measured, while venous blood flow was significantly slower at 1.5-7.1 cm/sec. Taking into account the corresponding vessel diameters ranging from 800 microm to 1.8 mm, blood flow rates of 3.0-26 ml/min in arteries and 1.2-4.8 ml/min in veins are obtained. The results were compared to ultrasound measurements, resulting in comparable blood flow velocities in the same subjects. Magn Reson Med 45:716-719, 2001.