Biocompatibility and performance of the Wallstent and several covered stents in a sheep iliac artery model
- PMID: 11287514
- DOI: 10.1016/s1051-0443(07)61916-2
Biocompatibility and performance of the Wallstent and several covered stents in a sheep iliac artery model
Abstract
Purpose: To compare the biocompatibility and performance of various stent-grafts to those of a bare stent in an ovine model with a subchronic (3 months) endpoint.
Materials and methods: Three different types of stent-grafts (ePTFE/nitinol, n = 8; polyester/nitinol, n = 8; and polycarbonate urethane/cobalt-alloy, n = 8) and a bare stent as a control (Ni-Co-Ti-steel-alloy, n = 8) were implanted in the iliac arteries in eight female sheep. One type of each stent-graft was implanted per animal, two implants at each side. The implantation sites for each type varied from animal to animal. Angiographic control and intravascular ultrasound (IVUS) imaging were performed before and after implantation, after 2 months, and before explantation at 3 months and were used to characterize patency and to assess intimal hyperplasia. After 3 months, the implants were retrieved and subjected to histologic evaluation (after methacrylate embedding, cutting, and histologic staining) to characterize the biologic response.
Results: Implantation was technically successful in all procedures. At 2 and 3 months after implantation, all segments in which stents had been implanted were patent. Marked neointima formation was found in the polyester-covered stent-graft that showed significant luminal narrowing of 50%, compared to the ePTFE-covered (24%) and polycarbonate urethane-covered endoprostheses (22%), as well as the bare stent (Wallstent; 9%; P < .001). A minimal inflammatory vessel wall reaction was demonstrated for the polyester-covered and ePTFE-covered endoprostheses; the polycarbonate urethane-covered stent-graft's response was demonstrable but not significantly different from that of the Wallstent. At 3 months, the ePTFE-covered stent-graft showed incomplete (>90%) endothelial coverage; in the other endoprostheses, complete but partially immature endothelialization was found.
Conclusion: All stent-grafts induced an inflammatory vessel wall reaction with neointimal hyperplasia. The polyester-covered endoprosthesis caused a marked reaction with 50% luminal stenosis. Endothelialization was retarded with the ePTFE-covered stent-graft. The bare stent performed best in regard to neointimal formation and caused the least inflammatory response.
Similar articles
-
Biocompatibility and performance of the Wallstent and the Wallgraft, Jostent, and Hemobahn stent-grafts in a sheep model.J Vasc Interv Radiol. 2002 Aug;13(8):823-30. doi: 10.1016/s1051-0443(07)61992-7. J Vasc Interv Radiol. 2002. PMID: 12171986
-
Histopathologic evaluation of an expanded polytetrafluoroethylene-nitinol stent endoprosthesis in canine iliofemoral arteries.J Vasc Interv Radiol. 1999 Apr;10(4):445-56. doi: 10.1016/s1051-0443(99)70064-3. J Vasc Interv Radiol. 1999. PMID: 10229474
-
Comparison of small intestinal submucosa-covered and noncovered nitinol stents in sheep iliac arteries: a pilot study.J Vasc Interv Radiol. 2002 May;13(5):489-98. doi: 10.1016/s1051-0443(07)61529-2. J Vasc Interv Radiol. 2002. PMID: 11997357
-
Evolving modalities for femoropopliteal interventions.J Endovasc Ther. 2009 Apr;16(2 Suppl 2):II82-97. doi: 10.1583/08-2654.1. J Endovasc Ther. 2009. PMID: 19624076 Review.
-
State of the Science in Tracheal Stents: A Scoping Review.Laryngoscope. 2022 Nov;132(11):2111-2123. doi: 10.1002/lary.29904. Epub 2021 Oct 15. Laryngoscope. 2022. PMID: 34652817
Cited by
-
Endovascular treatment of internal carotid and vertebral artery aneurysms using a novel pericardium covered stent.Interv Neuroradiol. 2012 Jun;18(2):164-71. doi: 10.1177/159101991201800207. Epub 2012 Jun 4. Interv Neuroradiol. 2012. PMID: 22681731 Free PMC article.
-
"Extarterectomy" for tumors invading the carotid artery: an animal model.Skull Base. 2007 Jul;17(4):229-34. doi: 10.1055/s-2007-984485. Skull Base. 2007. PMID: 18174922 Free PMC article.
-
Rapamycin-loaded nanoparticles for inhibition of neointimal hyperplasia in experimental vein grafts.J Cardiothorac Surg. 2011 May 12;6:69. doi: 10.1186/1749-8090-6-69. J Cardiothorac Surg. 2011. Retraction in: J Cardiothorac Surg. 2012 Mar 06;7:17. doi: 10.1186/1749-8090-7-17. PMID: 21569412 Free PMC article. Retracted.
-
Overlapped Stenting for Treatment of an Extracranial Carotid Artery Aneurysm.NMC Case Rep J. 2015 Jun 26;2(3):93-96. doi: 10.2176/nmccrj.2014-0432. eCollection 2015 Jul. NMC Case Rep J. 2015. PMID: 28663974 Free PMC article.
-
Assessment of Large Animal Vascular Dimensions for Intra-Aortic Device Research and Development: A Systematic Review.Innovations (Phila). 2023 Mar-Apr;18(2):144-151. doi: 10.1177/15569845231164134. Epub 2023 Apr 8. Innovations (Phila). 2023. PMID: 37029653 Free PMC article.