Reuse of healing abutments: an in vitro model of plasma cleaning and common sterilization techniques
- PMID: 11307410
Reuse of healing abutments: an in vitro model of plasma cleaning and common sterilization techniques
Abstract
The reuse of transgingival healing abutments has been advocated by several implant manufacturers, but cleaning and sterilization procedures to yield clean and optimal surfaces have yet to be developed. The objective of this in vitro project was to investigate various cleaning and sterilization regimens for the removal of biological debris to support reattachment of subgingival connective tissue. Simulated clinical healing abutment surfaces were exposed to culture medium with serum for 1 hour to simulate biological exposure. Simulated healing abutment surfaces not contaminated by serum were used to represent the "as-is" healing abutment surface without prior in vivo use. The discs were cleaned with detergent before sterilization by ultraviolet light (UV) or steam autoclaving (AC) both with and without 1- and 5-minute plasma cleaning (PC). A series of surface analytical techniques (XPS, AES, and surface contact angles) and in vitro analysis of cell attachment and spreading using gingival fibroblasts were performed. After exposure to the simulated biological conditions, clinical cleaning followed by UV resulted in contaminated surfaces and relatively high levels of cell attachment. PC before UV treatment enhanced surface energetics but did not affect cell attachment and spreading. AC increased surface wetting angles; which were decreased somewhat by previous PC. Cell attachment was significantly reduced by AC. Although some increase in cell attachment after longer plasma cleaning was noted in the AC group, no difference in cell spreading was seen in any AC group. Cell spreading seemed to be less for all AC groups compared with all UV, as-is, and control groups. Although certain cleaning (PC) and sterilization (UV) procedures can be effective for cleaning transgingival healing abutments, those using AC are questionable due to their propensity for organic and inorganic contamination and unfavorable surface alteration.
Similar articles
-
Effects of multiple sterilization on surface characteristics and in vitro biologic responses to titanium.J Oral Maxillofac Surg. 1996 Jun;54(6):738-46. doi: 10.1016/s0278-2391(96)90694-1. J Oral Maxillofac Surg. 1996. PMID: 8648479
-
In vitro periodontal ligament fibroblast attachment to plasma-cleaned titanium surfaces.J Oral Implantol. 1991;17(2):132-9. J Oral Implantol. 1991. PMID: 1811064
-
Short-term plasma-cleaning treatments enhance in vitro osteoblast attachment to titanium.J Oral Implantol. 1992;18(2):130-7. J Oral Implantol. 1992. PMID: 1289550
-
Should Healing Abutments and Cover Screws for Dental Implants be Reused? A Systematic Review.J Prosthodont. 2020 Jan;29(1):42-48. doi: 10.1111/jopr.13106. Epub 2019 Sep 16. J Prosthodont. 2020. PMID: 31453645
-
Biofilm on dental implants: a review of the literature.Int J Oral Maxillofac Implants. 2009 Jul-Aug;24(4):616-26. Int J Oral Maxillofac Implants. 2009. PMID: 19885401 Review.
Cited by
-
Evaluation of residual contamination on reused healing abutments.Clin Oral Investig. 2021 Oct;25(10):5889-5895. doi: 10.1007/s00784-021-03894-9. Epub 2021 Mar 25. Clin Oral Investig. 2021. PMID: 33763713
-
Effects of multiple implantations of titanium healing abutments: Surface characteristics and microbial colonization.Dent Mater. 2020 Sep;36(9):e279-e291. doi: 10.1016/j.dental.2020.05.016. Epub 2020 Jun 24. Dent Mater. 2020. PMID: 32591158 Free PMC article.
-
Soft tissue cell adhesion to titanium abutments after different cleaning procedures: preliminary results of a randomized clinical trial.Med Oral Patol Oral Cir Bucal. 2014 Mar 1;19(2):e177-83. doi: 10.4317/medoral.19329. Med Oral Patol Oral Cir Bucal. 2014. PMID: 24121917 Free PMC article. Clinical Trial.
-
Is the re-use of sterilized implant abutments safe enough? (Implant abutment safety).Med Oral Patol Oral Cir Bucal. 2019 Sep 1;24(5):e583-e587. doi: 10.4317/medoral.22967. Med Oral Patol Oral Cir Bucal. 2019. PMID: 31433387 Free PMC article.
-
The biocompatibility of titanium in a buffer solution: compared effects of a thin film of TiO2 deposited by MOCVD and of collagen deposited from a gel.J Mater Sci Mater Med. 2007 Oct;18(10):2075-83. doi: 10.1007/s10856-007-3133-3. Epub 2007 Jun 12. J Mater Sci Mater Med. 2007. PMID: 17562134