Novel, compact, intraoperative magnetic resonance imaging-guided system for conventional neurosurgical operating rooms
Abstract
Objective: Preliminary clinical experience with a novel, compact, intraoperative magnetic resonance imaging (MRI)-guided system that can be used in an ordinary operating room is presented.
Description of instrumentation: The system features an MRI scanner integrated with an optical and MRI tracking system. Scanning and navigation, which are operated by the surgeon, are controlled by an in-room computer workstation with a liquid crystal display screen. The scanner includes a 0.12-T permanent magnet with a 25-cm vertical gap, accommodating the patient's head. The field of view is 11 x 16 cm, encompassing the surgical area of interest. The magnet is mounted on a transportable gantry that can be positioned under the surgical table when not in use for scanning, thus rendering the surgical environment unmodified and allowing the use of standard instruments. The features of the integrated navigation system allow flap planning and intraoperative tracking based on updated images acquired during surgery.
Operative technique: Twenty patients with brain tumors were surgically treated using craniotomy or trans-sphenoidal approaches. One patient underwent conscious craniotomy with cortical mapping, and two underwent electrocorticography.
Experience and results: Planning was accurate. Resection control images were obtained for all patients during surgery, with precise localization of residual tumor tissue. There were no surgical complications related to the use of the system.
Conclusion: This intraoperative MRI system can function in a normal operating room modified only to eliminate radiofrequency interference. The operative environment is normal, and standard instruments can be used. The scanning and navigation capabilities of the system eliminate the inaccuracies that may result from brain shift. This novel type of intraoperative MRI system represents another step toward the introduction of the modality as a standard method in neurosurgery.
Similar articles
-
Glioma resection in a shared-resource magnetic resonance operating room after optimal image-guided frameless stereotactic resection.Neurosurgery. 2001 Apr;48(4):731-42; discussion 742-4. doi: 10.1097/00006123-200104000-00007. Neurosurgery. 2001. PMID: 11322433
-
Intraoperative magnetic resonance imaging with the magnetom open scanner: concepts, neurosurgical indications, and procedures: a preliminary report.Neurosurgery. 1998 Oct;43(4):739-47; discussion 747-8. doi: 10.1097/00006123-199810000-00005. Neurosurgery. 1998. PMID: 9766299
-
Serial intraoperative magnetic resonance imaging of brain shift.Neurosurgery. 2001 Apr;48(4):787-97; discussion 797-8. doi: 10.1097/00006123-200104000-00019. Neurosurgery. 2001. PMID: 11322439
-
Image-guided surgery of epilepsy.Neurosurg Clin N Am. 1996 Apr;7(2):229-43. Neurosurg Clin N Am. 1996. PMID: 8726438 Review.
-
Intra-operative magnetic resonance imaging in neurosurgery.Acta Neurochir (Wien). 2004 Jun;146(6):543-56; discussion 557. doi: 10.1007/s00701-004-0229-0. Epub 2004 May 21. Acta Neurochir (Wien). 2004. PMID: 15168222 Review.
Cited by
-
How to objectively evaluate the impact of image-guided surgery technologies.Eur J Nucl Med Mol Imaging. 2024 Aug;51(10):2869-2877. doi: 10.1007/s00259-023-06504-w. Epub 2023 Nov 16. Eur J Nucl Med Mol Imaging. 2024. PMID: 37971499 Review.
-
Origins of intraoperative MRI.Neurosurg Clin N Am. 2009 Apr;20(2):137-46. doi: 10.1016/j.nec.2009.04.002. Neurosurg Clin N Am. 2009. PMID: 19555875 Free PMC article.
-
Robotics in the neurosurgical treatment of glioma.Surg Neurol Int. 2015 Feb 13;6(Suppl 1):S1-8. doi: 10.4103/2152-7806.151321. eCollection 2015. Surg Neurol Int. 2015. PMID: 25722932 Free PMC article.
-
Low-field MRI: Clinical promise and challenges.J Magn Reson Imaging. 2023 Jan;57(1):25-44. doi: 10.1002/jmri.28408. Epub 2022 Sep 19. J Magn Reson Imaging. 2023. PMID: 36120962 Free PMC article. Review.
-
Dual-room 1.5-T intraoperative magnetic resonance imaging suite with a movable magnet: implementation and preliminary experience.Neurosurg Rev. 2012 Jan;35(1):95-109; discussion 109-10. doi: 10.1007/s10143-011-0336-3. Epub 2011 Jun 15. Neurosurg Rev. 2012. PMID: 21674146
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous