Interobserver and intraobserver reproducibility of peripheral blood and oxygen pressure measurements in the assessment of lower extremity arterial disease
- PMID: 11331846
- DOI: 10.1067/mva.2001.108011
Interobserver and intraobserver reproducibility of peripheral blood and oxygen pressure measurements in the assessment of lower extremity arterial disease
Abstract
Introduction: Peripheral blood pressure measurements play a prominent role in the diagnosis and follow-up of patients with peripheral vascular diseases. Toe pressure of the hallux (TP1) and second toe (TP2) and transcutaneous oxygen pressure (TCPO2) measurements are becoming more important. The ankle/brachial pressure index (ABPI) is known to be a reliable parameter, but the toe pressure and TCPO2 are evaluated less thoroughly. Therefore, we evaluated the reproducibility of TP1, TP2, TCPO2, ABPI, ankle pressure (AP), and brachial pressure (BP).
Patients and methods: In 54 patients with various stages of peripheral vascular disease, the intraobserver and interobserver reproducibility of BP, AP, ABPI, TP1, TP2, and TCPO2 was investigated by calculating the repeatability coefficient (RC) and the intraclass correlation coefficient (ICC) and by using Bland-Altman plots.
Results: The intraobserver and interobserver reproducibility at 1 day and after 1 week of BP, AP, ABPI, and TP1 was substantial and comparable (ICC range, 0.80-0.99), except for the BP after 1 week. The TP2 and TCPO2 were less reproducible (ICC range, 0.62-0.98). The interobserver RC of BP was 31 mm Hg; of AP, 44 mm Hg; of ABPI, 27%; of TP1, 41 mm Hg; of TP2, 67 mm Hg; and of TCPO2; 30 mm Hg. The difference plot showed that the observer variability was equally distributed across the range of pressure in all measurements.
Conclusion: The BP, AP, ABPI, and TP1 have a substantial intraobserver and interobserver reproducibility, whereas TP2 and TCPO2 show worse reproducibility. Especially when low values (or values around a cutoff value) are measured, the RC should be taken into account, and repetition of the measurement is advocated.
Similar articles
-
Reliability and repeatability of toe pressures measured with laser Doppler and portable and stationary photoplethysmography devices.Ann Vasc Surg. 2012 Apr;26(3):404-10. doi: 10.1016/j.avsg.2011.10.011. Epub 2012 Jan 30. Ann Vasc Surg. 2012. PMID: 22285350
-
A prospective comparison of bilateral photoplethysmography versus the ankle-brachial pressure index for detecting and quantifying lower limb peripheral arterial disease.J Vasc Surg. 2008 Apr;47(4):794-802. doi: 10.1016/j.jvs.2007.11.057. J Vasc Surg. 2008. PMID: 18381141
-
Transcutaneous oxygen and carbon dioxide pressure monitoring to determine severity of limb ischemia and to predict surgical outcome.J Vasc Surg. 1988 Apr;7(4):507-14. J Vasc Surg. 1988. PMID: 3352067
-
Clinical information content of transcutaneous oxymetry (tcpO2) in peripheral arterial occlusive disease (a review of the methodological and clinical literature with a special reference to critical limb ischaemia).Vasa. 1992;21(2):111-26. Vasa. 1992. PMID: 1621428 Review.
-
Investigation of arterial claudication with transcutaneous oxygen pressure at exercise: Interests and limits.Trends Cardiovasc Med. 2021 May;31(4):218-223. doi: 10.1016/j.tcm.2020.03.003. Epub 2020 Mar 14. Trends Cardiovasc Med. 2021. PMID: 32234333 Review.
Cited by
-
Toe-Brachial Index in the Second Toe: Substitutability to Toe-Brachial Index in the Great Toe and Ankle-Brachial Index.Ann Vasc Dis. 2016;9(4):300-306. doi: 10.3400/avd.oa.16-00078. Epub 2016 Nov 22. Ann Vasc Dis. 2016. PMID: 28018502 Free PMC article.
-
Risk factors and incidence for peripheral arterial disease in patients with typical lumbar spinal stenosis.Korean J Spine. 2014 Sep;11(3):183-7. doi: 10.14245/kjs.2014.11.3.183. Epub 2014 Sep 30. Korean J Spine. 2014. PMID: 25346766 Free PMC article.
-
Quality Control of the Foot Revascularization Using Indocyanine Green Fluorescence Imaging.World J Surg. 2017 Jul;41(7):1919-1926. doi: 10.1007/s00268-017-3950-6. World J Surg. 2017. PMID: 28265729
-
Evaluating the between-day reliability and likelihood of change of a test battery incorporating vastus lateralis muscle thickness, ankle-brachial pressure index, maximal voluntary torque, and six-minute walk test in patients with claudication.Vascular. 2025 Jun;33(3):687-694. doi: 10.1177/17085381241257735. Epub 2024 May 28. Vascular. 2025. PMID: 38807381 Free PMC article.
-
Effect of premeasurement rest time on systolic ankle pressure.J Am Heart Assoc. 2013 Aug 20;2(4):e000203. doi: 10.1161/JAHA.113.000203. J Am Heart Assoc. 2013. PMID: 23963754 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources