Comparison of 2-dimensional and 3-dimensional cardiac 82Rb PET studies
- PMID: 11337563
Comparison of 2-dimensional and 3-dimensional cardiac 82Rb PET studies
Abstract
Most new PET scanners have the capability to collect data in 3-dimensional (3D) (septa removed) mode. This allows many more detected events at the cost of increased random events and scatter. In the case of 82Rb imaging, the injected dose might have to be limited to avoid saturating the scanner. We present a comparison of 2-dimensional (2D) and 3D data collection for 82Rb cardiac studies using the ECAT EXACT scanner.
Methods: Resting 82Rb cardiac studies were collected in 2D and 3D modes for 33 consecutive patients. Four experienced physicians rated the images to determine if the different acquisition methods would lead to different patient care. A separate quantitative analysis was performed on data from multiple scans of a thoracic phantom filled to simulate cardiac and background radioactivity corresponding to 82Rb injections between 37 and 1740 MBQ:
Results: The 2D and 3D studies were significantly different, with the image quality being poorer in the 3D studies. The scanner collected data at near its maximal counting rate for either 1480-MBq 2D or 37-MBq 3D acquisitions. Because the data collection was counting rate limited in either mode, and there are more random and scatter events in 3D mode, the 2D acquisitions resulted in more detected true events and a better signal-to-noise ratio.
Conclusion: Cardiac 82Rb studies should be performed in 2D mode when using the ECAT EXACT scanner.
Similar articles
-
Comparison of 2-dimensional and 3-dimensional 82Rb myocardial perfusion PET imaging.J Nucl Med. 2003 Aug;44(8):1350-6. J Nucl Med. 2003. PMID: 12902427
-
A lesion detection observer study comparing 2-dimensional versus fully 3-dimensional whole-body PET imaging protocols.J Nucl Med. 2004 Apr;45(4):714-23. J Nucl Med. 2004. PMID: 15073270
-
[Reduction of radiation exposure in PET examinations by data acquisition in the 3D mode].Nuklearmedizin. 1999 Apr;38(3):75-9. Nuklearmedizin. 1999. PMID: 10320992 German.
-
Performance of list mode data acquisition with ECAT EXACT HR and ECAT EXACT HR+ positron emission scanners.Ann Nucl Med. 2006 Apr;20(3):189-94. doi: 10.1007/BF03027429. Ann Nucl Med. 2006. PMID: 16715949
-
PET imaging with yttrium-86: comparison of phantom measurements acquired with different PET scanners before and after applying background subtraction.Eur J Nucl Med Mol Imaging. 2003 May;30(5):716-20. doi: 10.1007/s00259-002-1112-y. Epub 2003 Feb 26. Eur J Nucl Med Mol Imaging. 2003. PMID: 12605273
Cited by
-
Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging.BMC Med Phys. 2009 Jan 29;9:2. doi: 10.1186/1756-6649-9-2. BMC Med Phys. 2009. PMID: 19178700 Free PMC article.
-
Myocardial perfusion imaging with PET.Imaging Med. 2013 Feb 1;5(1):35-46. doi: 10.2217/iim.13.1. Imaging Med. 2013. PMID: 23671459 Free PMC article.
-
Direct comparison between 2-dimensional and 3-dimensional PET acquisition modes for myocardial blood flow absolute quantification with O-15 water and N-13 ammonia.J Nucl Cardiol. 2006 Mar-Apr;13(2):220-4. doi: 10.1007/BF02971246. J Nucl Cardiol. 2006. PMID: 16580958
-
Topological Organization of Metabolic Brain Networks in Pre-Chemotherapy Cancer with Depression: A Resting-State PET Study.PLoS One. 2016 Nov 10;11(11):e0166049. doi: 10.1371/journal.pone.0166049. eCollection 2016. PLoS One. 2016. PMID: 27832148 Free PMC article.
-
Whole-body biodistribution and the influence of body activity on brain kinetic analysis of the 11C-PiB PET scan.Radiol Phys Technol. 2017 Dec;10(4):464-474. doi: 10.1007/s12194-017-0419-0. Epub 2017 Sep 11. Radiol Phys Technol. 2017. PMID: 28895034
Publication types
MeSH terms
Substances
LinkOut - more resources