Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling
- PMID: 11337581
Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling
Abstract
The purpose of this study was to investigate the quantitative properties and effects of ordered-subset expectation maximization (OSEM) on kinetic modeling compared with filtered backprojection (FBP) in dynamic PET studies. Both phantom and patient studies were performed.
Methods: For phantom studies dynamic two-dimensional emission scans with 10-min frames and 20-min scan intervals were acquired over a 14-h period using an HR+ PET scanner. Various phantoms were scanned: 2-, 5-, 10-, and 20-cm-diameter phantoms filled with an 18F solution (300 kBq/mL) and a NEMA phantom filled with an 18F background (40 kBq/mL) and a cold or 11C insert (450 kBq/mL). Transmission (Tx) scans of 5-60 min were acquired. Data were reconstructed using FBP Hanning 0.5 and OSEM with 2-12 iterations and 12 or 24 subsets. Quantitative accuracy and noise characteristics were assessed. For patient studies, five cardiac, three oncologic, and three brain dynamic 18F-FDG scans were used. Five reconstructions were performed: FBP Hanning 0.5, and OSEM 2 x 12 and OSEM 4 x 16 with and without 5-mm full width at half maximum smoothing. Time-activity curves were calculated using volumes of interest. The input function was derived from arterial sampling. Metabolic rate of glucose (MRglu) was calculated with a standard two-tissue compartment model and Patlak analysis.
Results: Contribution of Tx noise to the reconstructed image was smaller for OSEM than for FBP. Differences in signal-to-noise ratio between FBP and OSEM depended on number of iterations and phantom size. Bias with OSEM was observed for regions enclosed within a 5- to 10-fold hotter background. For cardiac studies OSEM 2 x 12 and OSEM 4 x 16 resulted in 13% and 21% higher pixel values and 9% and 15% higher MRglu values compared with FBP. Smoothing decreased all these values to 2%. Similar results were found for most tumor studies. For brain studies MRglu of FBP and OSEM 4 x 16 agreed within 2%. Use of OSEM image-derived input functions for cardiac PET studies resulted in a decrease in calculated MRglu of about 15%.
Conclusion: For most PET studies OSEM has equal quantitative accuracy as FBP. The higher pixel and MRglu values are explained by the better resolution of OSEM. However, OSEM does not provide accurate image-derived input functions for FDG cardiac PET studies because of bias in regions located within a hotter background.
Similar articles
-
Postinjection transmission scanning in myocardial 18F-FDG PET studies using both filtered backprojection and iterative reconstruction.J Nucl Med. 2004 Feb;45(2):169-75. J Nucl Med. 2004. PMID: 14960632 Clinical Trial.
-
Noise reduction in oncology FDG PET images by iterative reconstruction: a quantitative assessment.J Nucl Med. 2001 Sep;42(9):1316-23. J Nucl Med. 2001. PMID: 11535719
-
Quantitative comparison of analytic and iterative reconstruction methods in 2- and 3-dimensional dynamic cardiac 18F-FDG PET.J Nucl Med. 2004 Dec;45(12):2008-15. J Nucl Med. 2004. PMID: 15585474
-
Fast Statistical Iterative Reconstruction for Mega-voltage Computed Tomography.J Med Invest. 2020;67(1.2):30-39. doi: 10.2152/jmi.67.30. J Med Invest. 2020. PMID: 32378615 Review.
-
Four-dimensional (4D) image reconstruction strategies in dynamic PET: beyond conventional independent frame reconstruction.Med Phys. 2009 Aug;36(8):3654-70. doi: 10.1118/1.3160108. Med Phys. 2009. PMID: 19746799 Review.
Cited by
-
Construction and evaluation of quantitative small-animal PET probabilistic atlases for [¹⁸F]FDG and [¹⁸F]FECT functional mapping of the mouse brain.PLoS One. 2013 Jun 7;8(6):e65286. doi: 10.1371/journal.pone.0065286. Print 2013. PLoS One. 2013. PMID: 23762335 Free PMC article.
-
Blood flow and glucose metabolism in stage IV breast cancer: heterogeneity of response during chemotherapy.Mol Imaging Biol. 2008 Nov-Dec;10(6):356-63. doi: 10.1007/s11307-008-0163-2. Epub 2008 Aug 16. Mol Imaging Biol. 2008. PMID: 18709508 Free PMC article.
-
Direct parametric reconstruction in dynamic PET myocardial perfusion imaging: in vivo studies.Phys Med Biol. 2017 May 7;62(9):3539-3565. doi: 10.1088/1361-6560/aa6394. Epub 2017 Apr 5. Phys Med Biol. 2017. PMID: 28379843 Free PMC article.
-
An update on the use of image-derived input functions for human PET studies: new hopes or old illusions?EJNMMI Res. 2023 Nov 10;13(1):97. doi: 10.1186/s13550-023-01050-w. EJNMMI Res. 2023. PMID: 37947880 Free PMC article. Review.
-
Preclinical evaluation and quantification of [¹⁸F]MK-9470 as a radioligand for PET imaging of the type 1 cannabinoid receptor in rat brain.Eur J Nucl Med Mol Imaging. 2012 Sep;39(9):1467-77. doi: 10.1007/s00259-012-2163-3. Epub 2012 Jun 15. Eur J Nucl Med Mol Imaging. 2012. PMID: 22699528