Poorly fitting soft lenses affect ocular integrity
- PMID: 11352451
Poorly fitting soft lenses affect ocular integrity
Abstract
Purpose: The purpose of this study was to determine whether borderline fitting soft lenses have a greater effect on ocular physiology than well fitting lenses.
Method: The study was a retrospective analysis of data from seven soft lens clinical studies. The results of 63 eyes wearing borderline fitting lenses (31 tight and 32 loose fittings) were evaluated and compared with control eyes wearing optimally fitting lenses. Tight and loose lens fittings were defined as those showing tightness on push-up of > or =65% and < or =35%, respectively. An optimally fitting lens was defined as one showing tightness on push-up in the range 45-50% and decentration less than 0.7 mm. Slit lamp findings at the 1-month follow-up visit were compared.
Results: There was a significantly greater level of corneal staining among the tight lens wearing group than the controls (P=0.001); 68% of eyes wearing tight lenses showed staining as compared with 42% of eyes wearing well fitting lenses. There was also a greater level of conjunctival staining among tight lens wearing eyes but no significant difference in bulbar or limbal hyperemia. The loose lens wearing eyes showed a higher level of corneal staining: 69% vs 39% (P=0.04). In addition, the loose lens wearing eyes showed a higher level of bulbar (P=0.03) and limbal hyperemia (P=0.006). From supplementary analysis of 408 lens fittings, the incidence of corneal staining was found to increase with increasing tightness or looseness of fit.
Conclusion: The results show that corneal staining and other physiological effects of soft lens wear vary with the quality of lens fit suggesting that practitioners should be less tolerant of sub-optimal fitting soft lenses.
Similar articles
-
Prevalence of ocular surface symptoms, signs, and uncomfortable hours of wear in contact lens wearers: the effect of refitting with daily-wear silicone hydrogel lenses (senofilcon a).Eye Contact Lens. 2006 Dec;32(6):281-6. doi: 10.1097/01.icl.0000224522.04723.7a. Eye Contact Lens. 2006. PMID: 17099389
-
Daily versus monthly disposable contact lens: Which is better for ocular surface physiology and comfort?Cont Lens Anterior Eye. 2018 Jun;41(3):252-257. doi: 10.1016/j.clae.2017.12.005. Epub 2017 Dec 6. Cont Lens Anterior Eye. 2018. PMID: 29221708
-
Clinical comparison of optimum and large diameter soft contact lenses.Cont Lens Anterior Eye. 2018 Oct;41(5):405-411. doi: 10.1016/j.clae.2018.03.004. Epub 2018 May 5. Cont Lens Anterior Eye. 2018. PMID: 29739665 Clinical Trial.
-
Review of 20 years of soft contact lens wearer ocular physiology data.Cont Lens Anterior Eye. 2022 Feb;45(1):101525. doi: 10.1016/j.clae.2021.101525. Epub 2021 Oct 19. Cont Lens Anterior Eye. 2022. PMID: 34686430 Review.
-
Are hard lenses superior to soft? The advantages of soft lenses.Cornea. 1990;9 Suppl 1:S12-4; discussion S15. doi: 10.1097/00003226-199010001-00006. Cornea. 1990. PMID: 2189672 Review.
Cited by
-
Complications and compliance in professionally-managed and self-managed contact lenses compared with non-contact lens wearers.PLoS One. 2024 Sep 6;19(9):e0308538. doi: 10.1371/journal.pone.0308538. eCollection 2024. PLoS One. 2024. PMID: 39240979 Free PMC article.
-
Contact lenses, migraine, and allodynia.Oman J Ophthalmol. 2016 May-Aug;9(2):101-3. doi: 10.4103/0974-620X.184528. Oman J Ophthalmol. 2016. PMID: 27433037 Free PMC article.
-
Safety and Efficacy of a New Water Gradient Biomimetic Monthly Replacement Spherical Contact Lens Material (Lehfilcon A).Clin Ophthalmol. 2022 Aug 30;16:2873-2884. doi: 10.2147/OPTH.S362926. eCollection 2022. Clin Ophthalmol. 2022. PMID: 36065354 Free PMC article. Clinical Trial.
-
Characterization of Soft Contact Lens Fitting Using Ultra-Long Scan Depth Optical Coherence Tomography.J Ophthalmol. 2017;2017:5763172. doi: 10.1155/2017/5763172. Epub 2017 Jun 1. J Ophthalmol. 2017. PMID: 28751981 Free PMC article.
-
Micrometer-scale contact lens movements imaged by ultrahigh-resolution optical coherence tomography.Am J Ophthalmol. 2012 Feb;153(2):275-283.e1. doi: 10.1016/j.ajo.2011.06.023. Epub 2011 Sep 13. Am J Ophthalmol. 2012. PMID: 21920493 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources