Controlled comparison of original vented aerobic fan medium with new nonvented BacT/ALERT FA medium for culturing blood
- PMID: 11376041
- PMCID: PMC88095
- DOI: 10.1128/JCM.39.6.2098-2101.2001
Controlled comparison of original vented aerobic fan medium with new nonvented BacT/ALERT FA medium for culturing blood
Abstract
To evaluate the performance of BacT/ALERT FA (FA) medium, a new aerobic BacT/ALERT FAN (FAN) medium (Organon Teknika Corporation, Durham, N.C.) that does not require the added cost and inconvenience of a venting unit, we inoculated blood specimens from adult patients with suspected sepsis into an original FAN aerobic culture bottle and an FA bottle. Of 7,745 blood culture sets containing both bottles, 5,256 (68%) met the criteria for adequacy of filling. A total of 466 isolates judged to represent the causes of true infections were recovered from 276 patients; 271 isolates were recovered from both bottles, 82 were recovered from the FAN bottle only, and 113 were recovered from the FA bottle only (P < 0.05). More Burkholderia cepacia isolates (P < 0.01), Candida albicans isolates (P < 0.001), Cryptococcus neoformans isolates (P < 0.01), yeasts overall (P < 0.001), and total microorganisms (P < 0.05) were recovered from FA bottles. Of cultures found to be positive within the first 72 h of incubation, the mean times to detection were almost identical for FAN (20.4 h) and FA (20.7 h) bottles. Of 263 isolates that caused monomicrobic episodes of bloodstream infections, 180 were detected in both bottles, 32 were detected in FAN bottles only, and 51 were detected in FA bottles only (P < 0.05). Of 186 isolates considered to be contaminants, 63 were detected in both media, 64 were detected in FAN bottles only, and 59 were detected in FA bottles only (P was not significant). The number of false-positive results were comparable: 69 (1.3%) in FAN bottles and 56 (1.1%) in FA bottles. However, there were 14 isolates with false-negative results (6 yeasts, 6 nonfermenters, and 1 isolate each of Propionibacterium acnes and coagulase-negative staphylococci) in FAN bottles, whereas there were none in FA bottles. On the basis of these results, we conclude that the new nonvented FA bottle is superior to the original vented FAN medium for the recovery of B. cepacia and yeasts, especially C. albicans and C. neoformans, and is comparable to FAN medium for other microorganisms.
Similar articles
-
Comparison of the BacT/Alert PF pediatric FAN blood culture bottle with the standard pediatric blood culture bottle, the Pedi-BacT.J Clin Microbiol. 2001 Aug;39(8):2880-3. doi: 10.1128/JCM.39.8.2880-2883.2001. J Clin Microbiol. 2001. PMID: 11474007 Free PMC article.
-
Controlled clinical comparison of the BacT/ALERT FN and the standard anaerobic SN blood culture medium.J Clin Microbiol. 2004 Oct;42(10):4581-5. doi: 10.1128/JCM.42.10.4581-4585.2004. J Clin Microbiol. 2004. PMID: 15472312 Free PMC article.
-
Controlled comparison of BacT/ALERT FAN aerobic medium and BATEC fungal blood culture medium for detection of fungemia.J Clin Microbiol. 2001 Feb;39(2):622-4. doi: 10.1128/JCM.39.2.622-624.2001. J Clin Microbiol. 2001. PMID: 11158118 Free PMC article.
-
Laboratory detection of fungemia.Clin Lab Med. 1994 Mar;14(1):83-97. Clin Lab Med. 1994. PMID: 8181236 Review.
-
Rapid detection and identification of microorganisms from blood cultures.Clin Lab Med. 1994 Mar;14(1):181-95. Clin Lab Med. 1994. PMID: 8181231 Review.
Cited by
-
Evaluation of acridine orange staining as a replacement of subcultures for BacT/ALERT-positive, gram stain-negative blood cultures.J Clin Microbiol. 2003 Nov;41(11):5238-9. doi: 10.1128/JCM.41.11.5238-5239.2003. J Clin Microbiol. 2003. PMID: 14605172 Free PMC article.
-
Direct comparison of the BACTEC 9240 and BacT/ALERT 3D automated blood culture systems for candida growth detection.J Clin Microbiol. 2004 Jan;42(1):115-8. doi: 10.1128/JCM.42.1.115-118.2004. J Clin Microbiol. 2004. PMID: 14715740 Free PMC article.
-
Comparison of the BacT/Alert PF pediatric FAN blood culture bottle with the standard pediatric blood culture bottle, the Pedi-BacT.J Clin Microbiol. 2001 Aug;39(8):2880-3. doi: 10.1128/JCM.39.8.2880-2883.2001. J Clin Microbiol. 2001. PMID: 11474007 Free PMC article.
-
Comparison of BACTEC PLUS blood culture media to BacT/Alert FA blood culture media for detection of bacterial pathogens in samples containing therapeutic levels of antibiotics.J Clin Microbiol. 2007 Mar;45(3):816-21. doi: 10.1128/JCM.02064-06. Epub 2006 Dec 13. J Clin Microbiol. 2007. PMID: 17166960 Free PMC article.
-
The era of molecular and other non-culture-based methods in diagnosis of sepsis.Clin Microbiol Rev. 2010 Jan;23(1):235-51. doi: 10.1128/CMR.00043-09. Clin Microbiol Rev. 2010. PMID: 20065332 Free PMC article. Review.
References
-
- McNemar Q. Psychological statistics. 3rd ed. New York, N.Y: John Wiley & Sons, Inc.; 1962. pp. 209–239.
-
- Murray P R, Baron E J, Pfaller M A, Tenover F C, Yolken R H, editors. Manual of clinical microbiology. 7th ed. Washington, D.C.: American Society for Microbiology; 1999.
-
- Weinstein M P, Towns M L, Quartey S M, Mirrett S, Reimer L G, Parmigiani G, Reller L B. The clinical significance of positive blood cultures in the 1990s: a prospective comprehensive evaluation of the microbiology, epidemiology, and outcome of bacteremia and fungemia in adults. Clin Infect Dis. 1997;24:584–602. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous