Comparison of a polymer conjugate-enhanced enzyme immunoassay to ligase chain reaction for diagnosis of Chlamydia trachomatis in endocervical swabs
- PMID: 11376077
- PMCID: PMC88131
- DOI: 10.1128/JCM.39.6.2306-2307.2001
Comparison of a polymer conjugate-enhanced enzyme immunoassay to ligase chain reaction for diagnosis of Chlamydia trachomatis in endocervical swabs
Abstract
Two endocervical swabs from each of 1,123 women were collected into manufacturer-supplied transport tubes and tested for Chlamydia trachomatis by a polymer conjugate-enhanced (PCE) enzyme immunoassay (EIA) (IDEIA PCE Chlamydia; DAKO) and a ligase chain reaction assay (LCx Chlamydia; Abbott). After confirmation by the EIA blocking test, the sensitivity of the IDEIA PCE remained at 91.8% and the specificity increased from 98.2 to 99.8% compared to LCx.
References
-
- Chernesky M A, Lee H, Schachter J, Burczak J D, Stamm W E, McCormack W M, Quinn T C. Diagnosis of Chlamydia trachomatis urethral infection in symptomatic and asymptomatic men by testing first-void urine in a ligase chain reaction assay. J Infect Dis. 1994;170:1308–1311. - PubMed
-
- Clarke L M, Sierra M F, Daidone B J, Lopez N, Covino J M, McCormack W M. Comparison of the Syva MicroTrak enzyme immunoassay and Gen-Probe PACE 2 with cell culture for diagnosis of cervical Chlamydia trachomatis infection in a high-prevalence female population. J Clin Microbiol. 1993;31:968–971. - PMC - PubMed