Clinical usefulness of a multielectrode basket catheter for idiopathic ventricular tachycardia originating from right ventricular outflow tract
Abstract
Introduction: It often is difficult to determine the optimal ablation site for idiopathic ventricular tachycardia (VT) originating from the right ventricular outflow tract (RVOT) when the VT or premature ventricular complex (PVC) does not occur frequently. The aim of our study was to evaluate the usefulness of a multielectrode basket catheter for ablation of idiopathic VT originating from the RVOT.
Methods and results: Radiofrequency (RF) catheter ablation was performed using a 4-mm tip, quadripolar catheter in 50 consecutive patients with 81 VTs originating from the RVOT with (basket group = 25 patients with 45 VTs) or without (control group = 25 patients with 36 VTs) predeployment of a multielectrode basket catheter composed of 64 electrodes. Deployment of the multielectrode basket catheter was possible and safe in all 25 patients in the basket group. Ablation was successful in 25 (100%) of 25 patients in the basket group and in 22 (88%) of 25 patients in the control group. The total number of RF applications and the number of RF applications per PVC morphology did not differ between the two groups. However, both the fluoroscopic and ablation procedure times per PVC morphology were shorter in the basket group than in the control group (36.8+/-14.1 min vs 52.0+/-32.5 min, P = 0.04; 60.0+/-14.6 vs 81.5+/-51.2 min, P = 0.05). This difference was more pronounced in the 29 patients in whom VT or PVC was not frequently observed.
Conclusion: The multielectrode basket catheter is safe and useful for determining the optimal ablation site in patients with idiopathic VT originating from the RVOT, especially in those without frequent VT or PVC.
Comment in
-
Electrospatial mapping technologies: here to stay or gone tomorrow?J Cardiovasc Electrophysiol. 2001 May;12(5):518-20. doi: 10.1046/j.1540-8167.2001.00518.x. J Cardiovasc Electrophysiol. 2001. PMID: 11386510 No abstract available.
Similar articles
-
Localization of precise origin of idiopathic ventricular tachycardia from the right ventricular outflow tract by a 12-lead ECG: a study of pace mapping using a multielectrode "basket" catheter.Pacing Clin Electrophysiol. 1999 Dec;22(12):1760-8. doi: 10.1111/j.1540-8159.1999.tb00408.x. Pacing Clin Electrophysiol. 1999. PMID: 10642129
-
Magnetic versus manual catheter navigation for mapping and ablation of right ventricular outflow tract ventricular arrhythmias: a randomized controlled study.Heart Rhythm. 2013 Aug;10(8):1178-83. doi: 10.1016/j.hrthm.2013.05.012. Epub 2013 May 18. Heart Rhythm. 2013. PMID: 23692891 Clinical Trial.
-
New simplified technique for 3D mapping and ablation of right ventricular outflow tract tachycardia.Pacing Clin Electrophysiol. 2005 May;28(5):397-403. doi: 10.1111/j.1540-8159.2005.09547.x. Pacing Clin Electrophysiol. 2005. PMID: 15869671
-
Left ventricular outflow tract tachycardia including ventricular tachycardia from the aortic cusps and epicardial ventricular tachycardia.Herz. 2007 May;32(3):226-32. doi: 10.1007/s00059-007-2977-0. Herz. 2007. PMID: 17497256 Review.
-
An organized approach to the localization, mapping, and ablation of outflow tract ventricular arrhythmias.J Cardiovasc Electrophysiol. 2013 Oct;24(10):1189-97. doi: 10.1111/jce.12237. Epub 2013 Sep 9. J Cardiovasc Electrophysiol. 2013. PMID: 24015911 Review.
Cited by
-
Novel algorithm for accelerated electroanatomic mapping and prediction of earliest activation of focal cardiac arrhythmias using mathematical optimization.Heart Rhythm. 2017 Jun;14(6):875-882. doi: 10.1016/j.hrthm.2017.03.001. Epub 2017 Mar 6. Heart Rhythm. 2017. PMID: 28279745 Free PMC article.
-
Radiofrequency catheter ablation of idiopathic right ventricular outflow tract arrhythmias.Indian Pacing Electrophysiol J. 2013 Jan;13(1):14-33. doi: 10.1016/s0972-6292(16)30585-x. Epub 2013 Jan 1. Indian Pacing Electrophysiol J. 2013. PMID: 23329871 Free PMC article.