Simulation of the radon and radon progeny detection by silicon photodiodes
- PMID: 11388730
- DOI: 10.1097/00004032-200106000-00011
Simulation of the radon and radon progeny detection by silicon photodiodes
Abstract
The use of silicon photodiodes for field radon measurements requires the utilization of measuring cells for protection against environmental electromagnetic fields. It is very important to study and optimize the impact of geometry conditions (dimensions of the measuring cell) on the photodiode detection efficiency that can be reduced more than 10-30 times. Two models (for volume-distributed nuclides around the detector and for radon progeny deposited on the diode surface radon progeny) were applied for simulation of the photodiode detection in measuring cells of different sizes. Their use allows an optimal choice of the most appropriate cell regarding the measurement conditions. Thus, for soil gas measurements cells we use a cell of 3.6 cm in diameter by 20 cm in height; for determination of the radon exhalation rate we use a cell of 16 cm in diameter by 14 cm in height; and for measurements in dwellings and large spaces we use cells larger than 20 cm and lower than 4 cm.
Similar articles
-
The use of silicon photodiodes for radon and progeny measurements.Health Phys. 2001 Jun;80(6):590-6. doi: 10.1097/00004032-200106000-00010. Health Phys. 2001. PMID: 11388729
-
Four passive sampling elements (quatrefoil)--I. Monitoring radon and its progeny by surface-contamination monitors.Radiat Prot Dosimetry. 2011 May;145(2-3):280-3. doi: 10.1093/rpd/ncr049. Epub 2011 Apr 11. Radiat Prot Dosimetry. 2011. PMID: 21486828
-
A new device to measure the activity size distribution of radon progeny in a low level environment.Health Phys. 1997 Sep;73(3):494-7. doi: 10.1097/00004032-199709000-00008. Health Phys. 1997. PMID: 9287091
-
Procedures for the determination of 222Rn exhalation and effective 226Ra activity in soil samples.Appl Radiat Isot. 1999 Jun;50(6):1039-47. doi: 10.1016/s0969-8043(98)00121-3. Appl Radiat Isot. 1999. PMID: 10355105
-
State-of-the-art instruments for measuring radon/thoron and their progeny in dwellings--a review.Health Phys. 1996 Dec;71(6):973-4. Health Phys. 1996. PMID: 8919087 No abstract available.