. 1979 Apr;34(2):166-76.
doi: 10.1136/thx.34.2.166.

Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis

Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis

N D Broom et al. Thorax. 1979 Apr.

Abstract

To maintain optimum mechanical properties in glutaraldehyde-treated heart-valve tissue the full collagen crimp geometry originally present in the relaxed fresh tissue should be retained. By varying the pressure at which glutaraldehyde fixation is carried out, considerable alterations to this crimp geometry can be achieved. The mechanical stiffness of the preserved tissue is consequently affected, and this in turn has a striking influence on both the opening behaviour of the valve and the degree of strain localisation in the leaflet tissue. A pressure of 100 mmHg eliminated the collagen crimp geometry entirely, and this resulted in the formation of sites of local strain or kinks in the valve leaflets during opening. It is expected that this strain localisation phenomenon will influence the long-term fatigue durability of the treated tissue. Pressures even as low as 4 mmHg result in significant reductions of crimp geometry. Fresh valves should therefore be fixed under a positive head of pressure sufficient only to ensure that the leaflets seal along their coapting free margins. A pressure of less than 1 mmHg was sufficient to achieve this. Leaflets of the commercially available Hancock valve show features similar to valves fixed in glutaraldehyde at about 100 mmHg pressure.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Connect Tissue Res. 1978;6(1):37-50 - PubMed
    1. J Thorac Cardiovasc Surg. 1978 Jan;75(1):121-30 - PubMed
    1. J Thorac Cardiovasc Surg. 1977 Aug;74(2):317-21 - PubMed
    1. J Biomech. 1977;10(11/12):707-24 - PubMed
    1. J Thorac Cardiovasc Surg. 1978 Apr;75(4):606-11 - PubMed

LinkOut - more resources