Fourth-generation enzyme-linked immunosorbent assay for the simultaneous detection of human immunodeficiency virus antigen and antibody
- PMID: 11427563
- PMCID: PMC88179
- DOI: 10.1128/JCM.39.7.2518-2524.2001
Fourth-generation enzyme-linked immunosorbent assay for the simultaneous detection of human immunodeficiency virus antigen and antibody
Abstract
The VIDAS HIV DUO Ultra, a fourth-generation immunoassay under development for the simultaneous detection of human immunodeficiency virus type 1 (HIV-1) p24 antigen and antibodies to HIV-1 and HIV-2, was evaluated. The enzyme-linked fluorescence immunoassay, performed on the automated VIDAS instrument, is claimed to detect early and established HIV infection. The assay was challenged with a total of 2,847 samples that included 74 members of 10 seroconversion panels, 9 p24 antigen-only-reactive members of a panel of group M clades, 503 consecutively collected samples from individuals seeking care in the University of Maryland Medical System, 1,010 samples from U.S. blood donors, 1,141 samples from patients in a high-incidence population in Trinidad, 83 samples from a clinic for sexually transmitted diseases in the Bahamas, 10 confirmed HIV-1 group O samples, and 16 confirmed HIV-2 samples from the Cote d'Ivoire. Reference tests were U.S. Food and Drug Administration-licensed HIV antibody screening, p24 antigen tests, HIV confirmatory assays, and the Roche Diagnostics Amplicor HIV-1 Monitor. The VIDAS HIV DUO Ultra demonstrated 100% sensitivity and 99.5% specificity overall, with a 99.7% specificity in low-risk individuals. The analytical sensitivity, as assessed by seroconversion panels and p24 antigen in samples, was equivalent to the sensitivity of the reference assays used to characterize these panels. The VIDAS HIV DUO Ultra is accurate, offers potential advantages over conventional HIV testing for time and cost savings, has walk-away capability, and correctly identifies both early and established HIV infections.
Figures
References
-
- Alpert N L. VIDAS Immunoassay System. Clin Instrum Syst. 1993;12:1–5.
-
- Constantine N T. Supplemental tests for HIV-1 infection. In: Constantine N T, Callahan J D, Watts D M, editors. Retrovirology and retroviral testing. Boca Raton, Fla: CRC Press, Inc.; 1992. pp. 59–85.
-
- Constantine N T, van der Groen G, Belsey E, Heymann D, Tamashiro H. Sensitivity of HIV antibody assays as determined by seroconversion panels. AIDS. 1994;8:1715–1720. - PubMed
-
- Constantine N T, Zekeng L, Sangare A K, Gurtler L, Saville R, Anhary H, Ketema F. Diagnostic challenges for rapid human immunodeficiency virus assays. Performance using HIV-1 group O, HIV-1 group M, and HIV-2 samples. J Hum Virol. 1997;1:45–51. - PubMed
-
- Constantine N T. HIV antibody testing. In: Cohen P T, Sande M A, Volberding P A, editors. The AIDS knowledge base. 3rd ed. Philadelphia, Pa: Lippincott-Williams & Wilkins; 1999. pp. 105–112.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical