Initial results with an electrothermal bipolar vessel sealer
- PMID: 11443443
- DOI: 10.1007/s004640080025
Initial results with an electrothermal bipolar vessel sealer
Abstract
Background: An electrothermal bipolar vessel sealer (EBVS; Ligasure, Valleylab, Boulder, CO, USA) was developed as an alternative to suture ligatures, hemoclips, staplers, and ultrasonic coagulators for ligating vessels and tissue bundles. The EBVS seals vessels up to 7 mm in diameter by denaturing collagen and elastin within the vessel wall and surrounding connective tissue. This study is the first to determine the clinical efficacy and safety of this instrument and delineate its potential timesavings in both experimental (animal) and clinical scenarios.
Methods: A prospective review of the author's clinical experience with the EBVS in laparoscopic and open operations from October 1998 to March 2000 was performed. In addition, five Yorkshire domestic pigs underwent 150-cm small intestine resections (n = 10) using the EBVS (n = 5) and suture ligatures (n = 5). Measurements included time to complete intestinal resection, the number of applications per minute for each method, and the presence of postapplication bleeding. Statistical analysis was performed using Student's t-test.
Results: The EBVS was used in 98 cases (46 laparoscopic and 52 open) with a mean of 43 applications (range, 10-150 applications) per case. The operations included 53 colon and/or small bowel resections (54.1%), 24 fundoplications (24.5%), 12 gastric resections (12.2%), 3 splenectomies, 2 pancreatectomies, 1 adrenalectomy, 1 bilateral salpingo-oopherectomy, 1 pancreatic cyst-jejunostomy, and 1 vagotomy with gastrojejunostomy. In all these cases, the EBVS was intended to be the only means of vessel ligation. An alternative ligation technique was required for bleeding in only 13 (0.3%) of more than 4,200 applications of the EBVS. No postoperative hemorrhagic complications occurred. There was an estimated mean reduction in operative time of 39 min per open procedure, and a mean prolongation in operative time of 8 min per laparoscopic procedure when the EBVS was used in lieu of suture ligatures, hemoclips, staplers, or ultrasonic coagulators. In the animal model, the mean time for completion of the intestinal resection was 251.9 s for the EBVS and 702.0 s for ligatures (p < 0.001). The mean number of applications per minute was 7.6 for the EBVS and 1.8 for ligatures (p < 0.001). No postapplication bleeding was seen.
Conclusions: Initial clinical results from the use of EBVS in laparoscopic and open procedures demonstrate it to be safe and effective, reducing operative time in open procedures. Suture ligatures, ties, hemoclips, and other ligating techniques were used rarely (0.3%) after an application of the EBVS. In an experimental animal model, the EBVS was significantly faster and more efficient (more applications per minute) than ligatures for intestinal resection.
Similar articles
-
Electrothermal bipolar vessel ligation improves operative time during laparoscopic total proctocolectomy: a large single-center experience.Surg Endosc. 2016 Jul;30(7):2840-7. doi: 10.1007/s00464-015-4565-3. Epub 2015 Oct 28. Surg Endosc. 2016. PMID: 26511115
-
Safety and efficacy of the electrothermal bipolar vessel sealer in trauma.Injury. 2009 May;40(5):564-6. doi: 10.1016/j.injury.2008.01.024. Epub 2008 May 19. Injury. 2009. PMID: 18490017
-
Laparoscopic sigmoid and rectal resection using an electrothermal bipolar vessel sealing device.J Laparoendosc Adv Surg Tech A. 2007 Dec;17(6):719-22. doi: 10.1089/lap.2006.0238. J Laparoendosc Adv Surg Tech A. 2007. PMID: 18158799
-
Energy Sources for Laparoscopic Colorectal Surgery: Is One Better than the Others?J Laparoendosc Adv Surg Tech A. 2016 Apr;26(4):264-9. doi: 10.1089/lap.2016.0076. Epub 2016 Mar 17. J Laparoendosc Adv Surg Tech A. 2016. PMID: 26986017 Review.
-
Radiofrequency versus ultrasonic energy in laparoscopic colorectal surgery: a metaanalysis of operative time and blood loss.Surg Endosc. 2012 Oct;26(10):2917-24. doi: 10.1007/s00464-012-2285-5. Epub 2012 May 12. Surg Endosc. 2012. PMID: 22580873 Review.
Cited by
-
An advanced bipolar device helps reduce the rate of postoperative pancreatic fistula in laparoscopic gastrectomy for gastric cancer patients: a propensity score-matched analysis.Langenbecks Arch Surg. 2022 Dec;407(8):3479-3486. doi: 10.1007/s00423-022-02692-5. Epub 2022 Oct 1. Langenbecks Arch Surg. 2022. PMID: 36181517 Free PMC article.
-
Hepatic surgery using the Ligasure vessel sealing system.World J Surg. 2005 Jan;29(1):110-2. doi: 10.1007/s00268-004-7541-y. World J Surg. 2005. PMID: 15592913
-
Electrothermal bipolar vessel ligation improves operative time during laparoscopic total proctocolectomy: a large single-center experience.Surg Endosc. 2016 Jul;30(7):2840-7. doi: 10.1007/s00464-015-4565-3. Epub 2015 Oct 28. Surg Endosc. 2016. PMID: 26511115
-
Use of a vessel-sealing device compared with Jaboulay's technique for adult hydrocelectomy surgery: a retrospective cohort study.Updates Surg. 2024 Oct;76(6):2455-2460. doi: 10.1007/s13304-024-01838-y. Epub 2024 Apr 12. Updates Surg. 2024. PMID: 38609689
-
Implementation of a novel efficacy score to compare sealing and cutting devices in a porcine model.Surg Endosc. 2018 Feb;32(2):1002-1011. doi: 10.1007/s00464-017-5778-4. Epub 2017 Aug 24. Surg Endosc. 2018. PMID: 28840380
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources