Case Reports
. 2001 Jul;65(7):685-7.
doi: 10.1253/jcj.65.685.

Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing

Affiliations
Free article
Case Reports

Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing

T Washizuka et al. Jpn Circ J. 2001 Jul.
Free article

Abstract

This report describes the clinical management of 2 patients with ventricular fibrillation (VF) who received inappropriate shocks from an implantable cardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoidosis was confirmed as the underlying heart disease in 1 patient and idiopathic dilated cardiomyopathy in the other. Within 2 months after ICD implantation, both patients received several inappropriate shocks during sinus rhythm. Stored electrograms showed decreased R-wave amplitudes and increased T-wave amplitudes. The ICD sensed both R- and T-waves as ventricular activation, which met the rate criteria for VF treatment. Reprogramming the sensing threshold in association with administration of a drug to slow the heart rate decreased the incidence of the inappropriate shocks in both patients, but these palliative measures did not completely suppress the inappropriate shocks. To avoid T-wave oversensing, the repositioning or adding of a sensing lead is required. The potential risk of T-wave oversensing in ICD patients who have small R-wave amplitudes should be recognized.