Effect of femoral component offset on polyethylene wear in total hip arthroplasty
- PMID: 11451111
- DOI: 10.1097/00003086-200107000-00019
Effect of femoral component offset on polyethylene wear in total hip arthroplasty
Abstract
Seventeen staged, bilateral total hip arthroplasties performed in 17 patients were reviewed to compare side-to-side polyethylene wear. Implants used on both sides were similar except for implant offset: one hip in each patient was replaced using a femoral component having a standard implant offset, whereas the other side had a lateral offset implant. The mean followup was 5.70 years (range, 2-10.2 years) on the side with a standard femoral implant and 5.67 years (range, 2-9.7 years) on the side with a lateralized femoral component. The only statistically different parameter between the sides was the femoral component offset. All other parameters affecting polyethylene wear, such as period of followup, head size, head type, cup size, cup inclination, medialization of cup, and patient-related factors were similar on both sides. On the side with a standard femoral component, the mean actual prosthetic offset (determined by manufacturer's specifications) was 35.2 mm and the radiologic offset was 31.5 mm. On the side with a lateralized femoral component, the actual prosthetic offset was 42.5 mm and the radiologic offset was 40.1 mm. The difference in offsets between the sides was statistically significant. The mean preoperative offset of the femur was 38.8 mm. Regression analysis revealed that only femoral component offset and cup size correlated significantly with linear wear rate. On the side with a standard femoral component, the linear wear rate was 0.21 mm per year, whereas on the side with a lateralized femoral component, the linear wear rate was 0.10 mm per year. The differences in the linear wear rates were significant. Lateralization of the femoral component in this series more closely restored preoperative hip biomechanics and significantly decreased polyethylene wear.
Similar articles
-
Does using a polyethylene RM press-fit cup modify the preparation of the acetabulum and acetabular offset in primary hip arthroplasty?Orthop Traumatol Surg Res. 2017 Sep;103(5):669-674. doi: 10.1016/j.otsr.2017.03.026. Epub 2017 Jun 1. Orthop Traumatol Surg Res. 2017. PMID: 28579018
-
Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size.Clin Orthop Relat Res. 2016 Feb;474(2):365-71. doi: 10.1007/s11999-015-4319-5. Clin Orthop Relat Res. 2016. PMID: 25903943 Free PMC article.
-
Prediction of Polyethylene Wear Rates from Gait Biomechanics and Implant Positioning in Total Hip Replacement.Clin Orthop Relat Res. 2017 Aug;475(8):2027-2042. doi: 10.1007/s11999-017-5293-x. Epub 2017 Mar 2. Clin Orthop Relat Res. 2017. PMID: 28255945 Free PMC article.
-
Femoral offset: anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty.Orthop Traumatol Surg Res. 2009 May;95(3):210-9. doi: 10.1016/j.otsr.2009.03.010. Epub 2009 May 6. Orthop Traumatol Surg Res. 2009. PMID: 19423418 Review.
-
New polymer materials in total hip arthroplasty. Evaluation with radiostereometry, bone densitometry, radiography and clinical parameters.Acta Orthop Suppl. 2005 Feb;76(315):3-82. Acta Orthop Suppl. 2005. PMID: 15790289 Review.
Cited by
-
[Templating and radiological outcome monitoring for elective total hip arthroplasty. Applied quality management principles for safe patient care].Unfallchirurg. 2011 Sep;114(9):776-85. doi: 10.1007/s00113-011-2029-3. Unfallchirurg. 2011. PMID: 21870133 German.
-
Mid-Term Outcomes of a Short Modular Neck-Preserving Cementless Hip Stem: A Retrospective Study With a 6-Year Minimum Follow-Up.Arthroplast Today. 2024 Apr 27;27:101387. doi: 10.1016/j.artd.2024.101387. eCollection 2024 Jun. Arthroplast Today. 2024. PMID: 38707589 Free PMC article.
-
Robotic-assisted compared with conventional total hip arthroplasty: systematic review and meta-analysis.Postgrad Med J. 2018 Jun;94(1112):335-341. doi: 10.1136/postgradmedj-2017-135352. Epub 2018 May 18. Postgrad Med J. 2018. PMID: 29776983 Free PMC article.
-
Blood metal ions after hybrid metal-on-polyethylene Exeter-Trident total hip replacement.J Orthop Traumatol. 2016 Jun;17(2):149-53. doi: 10.1007/s10195-015-0369-4. Epub 2015 Jul 24. J Orthop Traumatol. 2016. PMID: 26206329 Free PMC article.
-
The effect of femoral offset modification on gait after total hip arthroplasty.Acta Orthop. 2014 Apr;85(2):123-7. doi: 10.3109/17453674.2014.889980. Epub 2014 Feb 25. Acta Orthop. 2014. PMID: 24564749 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources