Comparative Study
. 2001 Jun;2(6):449-55.

Comparison of defibrillation thresholds using monodirectional electrical vector versus bidirectional electrical vector

Affiliations
  • PMID: 11453582
Comparative Study

Comparison of defibrillation thresholds using monodirectional electrical vector versus bidirectional electrical vector

L Libero et al. Ital Heart J. 2001 Jun.

Abstract

Background: Currently, two main lead configurations are used for implantable cardioverter-defibrillators (ICD). One generates a monodirectional electrical vector by using the can surface as an active part (hot can) together with a right ventricular defibrillation coil. The other one (TRIAD) produces a bidirectional electrical vector by adding a proximal defibrillation electrode on the same lead. The purpose of this prospective study was to determine whether there is a difference between these configurations in terms of the acute defibrillation threshold (DFT). The secondary objective was to evaluate the possible sequential effect of successive arrhythmia induction and defibrillation shocks on the final DFT value.

Methods: In 44 patients (37 males, 7 females, mean age 59.18 +/- 12.05 years; mean ejection fraction 35.21 +/- 11.69%), a Hot Can Ventak family ICD (Guidant, St. Paul, MN, USA) was implanted in a left pectoral pocket. During the implant procedure, step-down to failure DFT testing was performed twice in each patient using the two different above-mentioned configurations: the bidirectional and the monodirectional. The first configuration to be tested was determined by a 1:1 randomization by center.

Results: The step-down DFT protocol was followed in 35 patients. The average DFT was 8.6 +/- 4.0 J for TRIAD and 10.4 +/- 4.3 J for the monodirectional (p = 0.009) lead configuration; this represents a 16.3% decrease in the DFT using a bidirectional configuration. Furthermore, no relationship between the final DFT and the number of ventricular fibrillation inductions and shocks received was observed, confirming the secondary objective.

Conclusions: Compared to the monodirectional electrical vector, the bidirectional electrical vector is clearly more beneficial for the patient.