Changes in intraocular pressure after laser in situ keratomileusis
- PMID: 11471999
- DOI: 10.3928/1081-597X-20010701-03
Changes in intraocular pressure after laser in situ keratomileusis
Abstract
Purpose: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia and myopic astigmatism, and to assess the accuracy of Goldmann applanation tonometry measurements after LASIK in these eyes.
Methods: LASIK was performed on 166 eyes of 93 patients for correction of myopia and myopic astigmatism. Intraocular pressure was measured by Goldmann applanation tonometry at the central and temporal parts of the cornea before and at 1, 3, 6, and 12 months after LASIK. The amount of change in IOP was computed and its relation to different variables was evaluated by regression analysis.
Results: Intraocular pressure measured at the center of the cornea was reduced by a mean of 3.69 +/- 1.63 mmHg after LASIK. Multiple regression analysis showed that the decrease in IOP was related to the preoperative IOP and the change in central corneal thickness after LASIK. Measurements of IOP at the temporal part of the cornea were also reduced by a mean of 2.39 +/- 1.71 mmHg. There was wide variability in the amount of difference between the temporal and central measurements after LASIK (temporal measurements were higher than central by 0 to +4 mmHg).
Conclusion: LASIK for myopia produced underestimation of IOP measured by Goldmann applanation tonometry at the central part of the cornea by a mean of 3.69 +/- 1.63 mmHg. The decrease of IOP was related to preoperative IOP and the change in central corneal thickness after LASIK. Temporal Goldmann applanation tonometry measurements, although decreased after LASIK, were less reliable.
Similar articles
-
Measurement of intraocular pressure in LASIK and LASEK patients using the Reichert Ocular Response Analyzer and Goldmann applanation tonometry.J Refract Surg. 2008 Apr;24(4):366-70. doi: 10.3928/1081597X-20080401-09. J Refract Surg. 2008. PMID: 18500086
-
Effect of mild to moderate myopic correction by laser-assisted in situ keratomileusis on intraocular pressure measurements with goldmann applanation tonometer, tono-pen, and pneumatonometer.J Glaucoma. 2002 Dec;11(6):493-6. doi: 10.1097/00061198-200212000-00007. J Glaucoma. 2002. PMID: 12483093
-
A correction formula for the real intraocular pressure after LASIK for the correction of myopic astigmatism.J Refract Surg. 2006 Mar;22(3):263-7. doi: 10.3928/1081-597X-20060301-11. J Refract Surg. 2006. PMID: 16602315 Clinical Trial.
-
Interface fluid after laser in situ keratomileusis.J Refract Surg. 2003 Jul-Aug;19(4):455-9. doi: 10.3928/1081-597X-20030701-13. J Refract Surg. 2003. PMID: 12899478 Review.
-
Late-onset interface fluid syndrome: A case report and literature review.Semin Ophthalmol. 2022 Oct-Nov;37(7-8):839-848. doi: 10.1080/08820538.2022.2102928. Epub 2022 Jul 21. Semin Ophthalmol. 2022. PMID: 35866219 Review.
Cited by
-
Reliability of Intraocular Pressure Measurement by Goldmann Applanation Tonometry After Refractive Surgery: A Review of Different Correction Formulas.Clin Ophthalmol. 2020 Sep 24;14:2783-2788. doi: 10.2147/OPTH.S263856. eCollection 2020. Clin Ophthalmol. 2020. PMID: 33061262 Free PMC article. Review.
-
[Alterations in intraocular pressure and the CorVis parameters after LASIK].Ophthalmologe. 2017 May;114(5):445-449. doi: 10.1007/s00347-016-0366-x. Ophthalmologe. 2017. PMID: 27620918 German.
-
[Mobile intraocular pressure measurement. From palpation to initial clinical experience with the handheld dynamic contour tonometer].Ophthalmologe. 2010 Jul;107(7):676-8, 680-2. doi: 10.1007/s00347-010-2175-y. Ophthalmologe. 2010. PMID: 20571809 German.
-
Goldmann and modified Goldmann tonometry measuring intraocular pressure changes in eyes which underwent myopic laser in situ Keratomileusis and photorefractive keratectomy.BMC Ophthalmol. 2022 Dec 20;22(1):503. doi: 10.1186/s12886-022-02741-z. BMC Ophthalmol. 2022. PMID: 36539706 Free PMC article.
-
Intraocular Pressure Calculation in Myopic Patients After Laser-Assisted In Situ Keratomileusis.Clin Ophthalmol. 2020 Feb 21;14:509-516. doi: 10.2147/OPTH.S239329. eCollection 2020. Clin Ophthalmol. 2020. PMID: 32109987 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources