Body water compartment measurements: a comparison of bioelectrical impedance analysis with tritium and sodium bromide dilution techniques
- PMID: 11478832
- DOI: 10.1054/clnu.2001.0398
Body water compartment measurements: a comparison of bioelectrical impedance analysis with tritium and sodium bromide dilution techniques
Abstract
Background and aims: This study was undertaken to assess the comparability of body water compartment estimates in healthy volunteers using single and dual frequency bioelectrical impedance analysis (BIA) with established reference methods of tritium and NaBr dilution.
Methods: Total body water (TBW) was estimated in 10 healthy volunteers using single frequency (50 kHz) BIA (Bodystat 1500), dual frequency (5 and 200 kHz) BIA (Bodystat Dualscan 2005) and tritium dilution. Extracellular water (ECW) was measured with dual frequency BIA and NaBr dilution. BIA was performed using distal tetrapolar electrodes in the supine position. Venous blood was sampled for measurement of background concentrations of tritium and NaBr using a beta counter and high performance liquid chromatography respectively. 10 ml tritiated water (3.7 MBq) and 50 ml 5% NaBr solution were then injected intravenously and blood samples taken from the opposite arm every 45 min for 4.5 h for estimation of concentrations of tritium and NaBr.
Results: There was good correlation (r(2)=0.76) between estimates of ECW using dual frequency BIA and NaBr dilution, with the former overestimating ECW by approximately 1 L. However, this difference varied systematically with body weight. Although TBW measurements obtained by single and dual frequency BIA correlated well with estimates using tritium dilution (r(2)=0.96 and 0.95 respectively), single frequency BIA underestimated TBW by approximately 1 L and dual frequency BIA by approximately 5 L compared to tritium dilution.
Conclusion: TBW measurements obtained using the single frequency BIA device were more accurate than those obtained using the dual frequency BIA device. Dual frequency BIA provided a reasonably accurate estimate of ECW.
Copyright 2001 Harcourt Publishers Ltd.
Similar articles
-
Altered body water distribution in subjects with juvenile rheumatoid arthritis and its effects on the measurement of water compartments from bioelectric impedance.Eur J Clin Nutr. 1996 Jun;50(6):335-9. Eur J Clin Nutr. 1996. PMID: 8793411
-
Reliability of bioelectrical impedance methods in detecting body fluids in elderly patients with congestive heart failure.Scand J Clin Lab Invest. 2006;66(1):19-30. doi: 10.1080/00365510500402158. Scand J Clin Lab Invest. 2006. PMID: 16464784
-
Body water distribution in severe obesity and its assessment from eight-polar bioelectrical impedance analysis.Eur J Clin Nutr. 2005 Feb;59(2):155-60. doi: 10.1038/sj.ejcn.1602049. Eur J Clin Nutr. 2005. PMID: 15340370
-
Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods.Med Eng Phys. 2008 Dec;30(10):1257-69. doi: 10.1016/j.medengphy.2008.06.009. Epub 2008 Aug 3. Med Eng Phys. 2008. PMID: 18676172 Review.
-
Bioelectrical impedance analysis--part I: review of principles and methods.Clin Nutr. 2004 Oct;23(5):1226-43. doi: 10.1016/j.clnu.2004.06.004. Clin Nutr. 2004. PMID: 15380917 Review.
Cited by
-
Physical activity, nutritional status, and dietary habits of students of a medical university.Sport Sci Health. 2016;12:261-267. doi: 10.1007/s11332-016-0285-x. Epub 2016 Jun 13. Sport Sci Health. 2016. PMID: 27478514 Free PMC article.
-
Prevalence of normal weight obesity in Switzerland: effect of various definitions.Eur J Nutr. 2008 Aug;47(5):251-7. doi: 10.1007/s00394-008-0719-6. Epub 2008 Jul 5. Eur J Nutr. 2008. PMID: 18604623
-
Can the Total Body Water and Total Fat Mass Predict Kidney Stone Recurrence in Overweight and Obese Patients?Cureus. 2024 Nov 10;16(11):e73367. doi: 10.7759/cureus.73367. eCollection 2024 Nov. Cureus. 2024. PMID: 39659337 Free PMC article.
-
Physical activity, sedentary time and gain in overall and central body fat: 7-year follow-up of the ProActive trial cohort.Int J Obes (Lond). 2015 Jan;39(1):142-8. doi: 10.1038/ijo.2014.66. Epub 2014 Apr 15. Int J Obes (Lond). 2015. PMID: 24732143 Free PMC article. Clinical Trial.
-
The theory and fundamentals of bioimpedance analysis in clinical status monitoring and diagnosis of diseases.Sensors (Basel). 2014 Jun 19;14(6):10895-928. doi: 10.3390/s140610895. Sensors (Basel). 2014. PMID: 24949644 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources