Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2001 Aug;36(8):1231-3.
doi: 10.1053/jpsu.2001.25771.

Technical assessment of porcine enteroenterostomy performed with ZEUS robotic technology

Affiliations
Comparative Study

Technical assessment of porcine enteroenterostomy performed with ZEUS robotic technology

C M Hollands et al. J Pediatr Surg. 2001 Aug.

Abstract

Background/purpose: Limitations of minimally invasive pediatric surgery include the inability to perform precise anastomoses of 2 to 15 mm. Robotic technology facilitates the performance of endoscopic microsurgical procedures. This study examined the technical feasibility of performing an enteroenterostomy in piglets utilizing ZEUS robotic technology.

Methods: Ten piglets (6.5 to 8.5 kg) underwent enteroenterostomy. Standard laparoscopic techniques were used in the control group (n = 5), and ZEUS robotic technology was used in the experimental group (n = 5). AESOP controlled the camera in both groups. Anesthesia time; surgery time; robotic set-up time; and anastomotic time, patency, diameter, and integrity were compared.

Results: No statistical difference existed between the means of the control and experimental groups for anesthesia time (176.0 v 154.0 minute; P =.63), surgery time (143.0 v 139.2 minute; P =.92), anastomosis time (109.4 v 93.0 minutes; P =.56), AESOP set-up time (4.2 v 7.0 minutes; P =.51), and anastomotic diameter (7.062 v 7.362 mm; P =.62). All anastomoses were patent without narrowing. The ZEUS cases averaged 14 minutes faster than the standard laparoscopic cases, even with the ZEUS set-up time included.

Conclusions: These data supports the hypothesis that robotic-assisted enteroenterostomy is technically feasible. ZEUS robotic technology will potentially play an important role in expanding the applications of minimally invasive pediatric surgery.

PubMed Disclaimer

Publication types

LinkOut - more resources